Slot Commands - Quantifi Photonics SLED 1000 Series User Manual

Superluminescent led light source
Table of Contents

Advertisement

9.7.2  Slot commands
Command
:SLOT<slot>:OPC?
Syntax
:SLOT<slot>:OPC?
Description
Query the Operation Complete Status of the module
Parameters
N/A
Response
1: the module is ready to accept a new command
0: the module is busy performing a previous operation
NOTE: Any commands sent to the module when :MODUle<slot>:OPC? is NOT 1, may not execute or return an
error.
Syntax
:SLOT1:OPC?
-> 1
Command
:SLOT<slot>:OPTions?
Syntax
:SLOT<slot>:OPTions?
Description
Query the modules managed by the CohesionSCPI service
Parameters
N/A
Response
A comma separated array, or a single integer value based on the arguments given
Example
:SLOT3:OPTions?
-> 1,,,,,,,
Command
:SLOT<slot>:IDN?
Syntax
:SLOT<slot>:IDN?
Description
Query the slot identification
Parameters
N/A
Response
A comma-separated string containing "<manufacturer>,<model name>,<serial number>,<hardware
version><firmware version>".
Note that the hardware and firmware versions are not comma separated.
Example
:SLOT3:IDN? ->
Quantifi Photonics Ltd,SLED-1011-2-FA-PXIE,QP-192001,HW1.0FW1.00
Quantifi Photonics | SLED 1000 / 1300 Series PXIe User Manual | Document version 1.08
Summary >>
>>
 
 
 
 
 
 
Summary >>
>>
 
 
 
 
Summary >>
>>
 
 
 
 
45 / 61

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sled 1300 series

Table of Contents