for the inlet and outlet are obviously open to the elements. Our inlet and outlet ports are
protected by custom made fittings with dual o-ring seals for backup protection. As long as
the inlet and outlet sample lines are installed in a protected way as described earlier, then
the device is robust and protected from the elements. For continuous monitoring outdoors
we still recommend the instrument be further protected from the sun and precipitation. The
MA350 comes with included mounting hardware for wall mounting. AethLabs also offers a
pole mounting kit.
Temperature:
microAeth instruments are specified to operate between 5 ~ 40 °C. Furthermore, rapid
changes in temperature can impact data. Such effects are inversely propertional to
timebase and flow rate settings. It is possible that an environmentally controlled enclosure
could be helpful and/or necessary depending on the application.
Humidity:
As with many other types of particle measurement instrumentation, changes in relative
humidity can change the size and optical properties of particles. For locations with high
RH variability, an external sample dryer or heater may be used to condition the sample.
AethLabs offers a portable aerosol dryer for this application. Moving between a dry, air
conditioned environment, to a high humidity environment can impact the data while the
device equilibrates. Selecting a higher flowrate and longer timebase will help to reduce
these effects.
Contamination:
Preventative maintenance for your deployment and sampling apparatus are important
for protecting your instrument. Keeping the microAeth and its air passageways, internal
components, and optical chambers clean is critical for maintaining the instrument and
producing quality measurements. <bold> We recommend, at a minimum, Standard
Maintenance Service at AethLabs or an authorized service center on at least an annual
basis.
Typically, Black Carbon particles are smaller than one micron in diameter. In some
sampling conditions where the aerosol is primarily composed of light scattering particulate
matter, such as dust or smoke from biomass fuels, there can be a prevalent fraction
of larger-diameter particles. Contamination of the instrument can cause increased
measurement noise, poor sealing of the analytical area and degraded operational lifetime
of some components.
For these applications, and to add contamination protection for your microAeth in high
concentrations, AethLabs offers a microCyclone™ size selective inlet that will limit the size
of the particles entering the microAeth to less than 2.5 microns in diameter when used
at at the specified flow rate. The microCyclone may not be appropriate for use with the
microAeth MA200, MA300, or MA350 in some applications depending on required flow
rates for DualSpot® loading compensation or other application specific flow requirements.
Contact AethLabs for more information.
MA Series MA200, MA300, MA350 Quick Start Guide
microAeth
®
6
Recommendations and Best Use Practices
Instrument Settings: We recommend getting started using the 60 seconds timebase and
100mlpm flow rate settings, with 5 wavelengths enabled. After using the instrument in your
application and collecting some data, these settings can be easily changed depending on
your requirements.
SingleSpot™ vs DualSpot®: DualSpot® loading compensation method corrects for the
optical loading effect and provides additional information about aerosol optical properties.
This method simultaneously collects aerosol samples on two analysis spots in parallel and
at different face velocities while measuring the rate of change in absorption of transmitted
light due to the particles loading on the filter. The use of this method is an optional user
selectable feature in the microAeth Manager program. DualSpot is usually recommended,
though SingleSpot with a lowered ATN threshold of 50 (with UV enabled) usually has
minimal loading effect in the IR channel
The timebase, flow rate, and WiFi settings will have the greatest impact on battery run
time. NOTE: Battery life will start to gradually diminish after many cycles (~ 1 year of use).
Runtimes can vary based on individual MA series instruments.
Individual Data Point Noise:
Instrumental noise contributes a random perturbation to the 'ideal' BC data. Its magnitude
is inversely proportional to the operational parameters of timebase and flow rate. Data
collected on a 1 second timebase should always be smoothed or averaged over longer
periods, unless measuring high concentrations where the instrumental noise is negligible.
Effects of Contamination, Vibration, and Impact: Primarily affected by timebase setting.
Increased flow rate will also reduce the effect.
1 second
very large
Choosing the Tape Advance ATN for your application:
Choosing the right tape advance ATN trigger helps ensure the integrity of your
measurement by reducing possible loading effects. Loading effects are variable, aerosol
dependent, and are not always present. The range of the Tape Advance ATN trigger
is 1 to 100. The lower the Tape Advance ATN trigger setting, the lower the impact of
loading effects on the data at the expense of more tape consumption. Your settings from
SingleSpot™ vs DualSpot®, Flow Rate, and Wavelengths also impact how frequently you
reach your ATN setpoint and tape advance frequency.
MA Series MA200, MA300, MA350 Quick Start Guide
AethLabs
microAeth
®
5 seconds
10 seconds
30 seconds
large
large
moderate
60 seconds
300 seconds
low
least
AethLabs
7