Performing Measurements; To Start Sampling And Measrements; To Stop Sampling And Measrements; Usb Communication, Configuration, & Data Export - Aethlabs microAeth MA Series Quick Start Manual

Hide thumbs Also See for microAeth MA Series:
Table of Contents

Advertisement

Performing Measurements

The inlet and outlet ports are 10-32 UNF imperial (inch) threaded ports.
The inlet port, outlet port, and all connections to the inlet port must be properly
protected from the environment. There must be limited restriction to flow while
protection from water, insects, bugs, and other objects that can block or infiltrate the
instrument air pathway through both the inlet and outlet ports of the instrument. Extra
precaution must be taken as the internal pump of the instrument is pulling air into the
instrument through the inlet port.
inlet
outlet

To Start Sampling and Measrements:

1) Ensure there is a filter cartridge installed and clamped.
2) Confirm that all settings are correct for the sampling campaign.
3) To start measurement, scroll through menu options to 'Start Measurement' and press
the center button.
4) 'SAMPLER PREPARING' will be displayed on screen. The microAeth will automatically
advance the filter tape cartridge to a new sampling location and run its start up routine.
5) When sampling and measurement begins, the screen will display a "play"
the top left cornter, 'Press Center to Stop' on line 2, the DualSpot (DS) or SingleSpot (SS)
target total flow rate on line 3, and 'SAMPLING' on line 4.
MA Series MA200, MA300, MA350 Quick Start Guide
microAeth
®
20
inlet
outlet
outlet
inlet

To Stop Sampling and Measrements:

1) While the microAeth is sampling, the screen will display a "play"
left cornter, 'Press Center to Stop' on line 2, the DualSpot (DS) or SingleSpot (SS) target
total flow rate on line 3, and 'SAMPLING' on line 4.
2) Press and hold the center button for 2 seconds to stop sampling and measurement. The
screen will display a "stop"
displayed on the screen and sampling and measurement will finish.
USB Communication, Configuration, & Data Export
The microAeth instrument has a USB port for communication with the microAeth Manager
software for configuring the instrument settings and downloading data using the AethLabs
provided mini-B to Type A USB cable.
To connect and configure the microAeth using the microAeth Manager application:
1) Connect the microAeth instrument to your computer using the provided mini USB cable.
2) Turn on the microAeth instrument.
3) Open the microAeth Manager application. The microAeth information will be loaded into
the microAeth Manager instrument device pane.
symbol in
4)
Use the microAeth Manager to modify all available settings of the microAeth.
MA Series MA200, MA300, MA350 Quick Start Guide
AethLabs
microAeth
®
symbol in the top
symbol in the top left corner and 'STOPPED' will be
AethLabs
21