MA350 Included Items Operation The small size and light weight of the microAeth MA series instruments will allow you Turn On: to gather data in a wide range of mobile or stationary applications. The MA series filter...
Important information Before Getting Started 10) To get the best data from the microAeth MA Series instruments, we highly recommend that the instrument warm up for approximately 30 minutes so that it can 1) The microAeth Manager application for MacOS and Windows can be used to modify equilibrate.
RH variability, an external sample dryer or heater may be used to condition the sample. Runtimes can vary based on individual MA series instruments. AethLabs offers a portable aerosol dryer for this application. Moving between a dry, air Individual Data Point Noise: conditioned environment, to a high humidity environment can impact the data while the Instrumental noise contributes a random perturbation to the ‘ideal’...
10. Serial number label Cross-platform microAeth® Manager software Download (download via AethLabs website) Download Quick Start Guide (download via AethLabs website) Download Manual (download via AethLabs website) MA Series MA200, MA300, MA350 Quick Start Guide MA Series MA200, MA300, MA350 Quick Start Guide...
11. USB mini-B port (download via AethLabs website) 12. User interface buttons (3) Download Quick Start Guide (download via AethLabs website) 13. User interface screen 14. Flat head screw (1) in filter tape cartridge door Download Manual (download via AethLabs website) 15.
Press center button to select an Active Menu Option operation or setting to configure During Startup: The AethLabs logo is displayed followed by the Firmware Version, Serial Port Baud Rate, Instrument UTC Date / Time and AutoSample setting value. Measurement Status...
Press ‘OK’ to select current value and cartridge to remove the cartridge straight out of the instrument. return to the top menu level MA Series MA200, MA300, MA350 Quick Start Guide MA Series MA200, MA300, MA350 Quick Start Guide microAeth AethLabs microAeth AethLabs ® ®...
NONE) If reinstalling the same tape cartridge that was previously installed (no change to the tape cartridge was made - the tape cartridge was not removed after the tape clamp MA Series MA200, MA300, MA350 Quick Start Guide MA Series MA200, MA300, MA350 Quick Start Guide microAeth AethLabs microAeth AethLabs ® ®...
The microAeth instrument has a USB port for communication with the microAeth Manager inlet outlet inlet outlet software for configuring the instrument settings and downloading data using the AethLabs outlet inlet provided mini-B to Type A USB cable. To connect and configure the microAeth using the microAeth Manager application: 1) Connect the microAeth instrument to your computer using the provided mini USB cable.
Manage data window. MA Series MA200, MA300, MA350 Quick Start Guide MA Series MA200, MA300, MA350 Quick Start Guide microAeth AethLabs microAeth AethLabs ® ®...
If re-calibration is needed, it is possible to calibrate to within 1-2% on average using the AethLabs MA Series Flow Calibration Kit. Only the external mass flowmeter provided in Optical the AethLabs MA Series Flow Calibration Kit is compatible with the MA200, MA300, and Optical saturation saturation MA350.
Page 14
WiFi functionality. 4294967296 WiFi Line Full WiFi data buffer is full. Remote power Power down command received remotely via serial 274877906944 down connection. MA Series MA200, MA300, MA350 Quick Start Guide microAeth AethLabs ®...
Need help?
Do you have a question about the microAeth MA Series and is the answer not in the manual?
Questions and answers