Instrument Settings: Measurement Timebase and Flow Rate Longest Filter Life Shortest In order to get the best data from the microAeth AE51 for the sampling campaign, we 50 ml/min 100 ml/min 150 ml/min 200 ml/min highly recommend that the instrument warm up for approximately 10-15 minutes so that it can equilibrate to its environment.
PC Software Installation release button on the bottom. 1. Always make sure that a filter strip is 1. Do not connect the microAeth AE51 5. Loosen the rubber cover on the front of installed in the microAeth AE51 when to the computer until the software the AE51 by pulling the tab away from it is operating.
Page 4
1. Make sure that a filter strip is installed setup. in the microAeth AE51. Turn on the AE51 by depressing the power button for 4 seconds until the instrument beeps for the second time and the red and green LEDs illuminate together.
Page 5
1 second as follows: The time on the AE51 is displayed. In order to sync the time on the AE51 with 1. Turn on the microAeth AE51. 1. Press and hold the power button. the PC time, click the Sync with PC 2.
“dirty” instrument will create data excursions that are far larger than those of a “clean” • In order to get the best data from the microAeth AE51 for the sampling campaign, we unit. These effects are amplified greatly at the shorter timebase settings. Cleaning intervals highly recommend that the instrument warm up for approximately 10-15 minutes so that will vary based on the sampling environment and concentrations.
• Light source current too high or too low of 3 triple beeps. • Light source feedback circuit error. https://aethlabs.com San Francisco, California Quick Start Guide - microAeth AE51, Jul 2016, Rev 05 microAeth AE51 Quick Start Guide AethLabs ®...
Need help?
Do you have a question about the microAeth AE51 and is the answer not in the manual?
Questions and answers