Calibration

All microAeth instruments are flow calibrated at approximately standard atmospheric
pressure and room temperature to ensure accurate flow rates at the time of production.
Local operating conditions vary around the world. Temperature and pressure at the time of
calibration are recorded in the flow calibration table header. Before instrument deployment,
we suggest that the flow rates at the inlet of your microAeth instrument are checked.
In general, and especially for sampling at higher elevations, AethLabs suggests calibrating
the flow of your microAeth in the location at which you plan to use it. This will ensure that
the internal mass flowmeter is calibrated for the in situ atmospheric pressure.
If re-calibration is needed, it is possible to calibrate to within 1-2% on average using the
AethLabs MA Series Flow Calibration Kit. Only the external mass flowmeter provided in
the AethLabs MA Series Flow Calibration Kit is compatible with the MA200, MA300, and
MA350.
Cleaning & Maintenance
If the microAeth is exposed to any liquids or other damaging contaminants, immediately
turn off the instrument, disconnect all cables, and remove any foreign substances
in contact with the instrument. Do not use liquids or other cleaning products on the
instrument. Wait until the microAeth is completely dry before charging or turning on the
instrument. Only AethLabs authorized service personnel should clean the air passageways
and internal components of the microAeth.
AethLabs recommends sending your instrument for annual service, or more frequent
service depending on use and operating conditions. Cleaning intervals will vary based on
the sampling environment and concentrations.
In addition to any necessary repairs, AethLabs offers flow calibration service and standard
maintenance service (including full diagnostics, cleaning, flow calibration, and instrument
testing) at the AethLabs San Francisco office. AethLabs also has authorized repair
locations in Europe and South America.
MA Series MA200, MA300, MA350 Quick Start Guide
microAeth
®
24

Status Codes

Code Value
1
2
4
8
16
32
64
128
256
512
1024
2048
4096
8192
16384
MA Series MA200, MA300, MA350 Quick Start Guide
AethLabs
microAeth
®
Readable
Status Description
status
No status notifications
Start up
Instrument in sampling and measurement startup
Tape advance occurring during sampling and
Tape advance
measurement period
N/A
Optical
Optical saturation
saturation
Sample timing
Sample timing error
error
DualSpot
Sampling spot 2 of DualSpot
Enabled
compensation is active
Flow unstable during sampling and measurement
Flow unstable
period. Flow deviates from target flow setpoint by
more than ±5%.
Flow out of range during sampling and
Pump drive limit
measurement period
Time synchronization source is manual
Time source
(synchronized to application/computer time or no
manual
GPS time is available)
User skipped tape advance. No tape advance
occurred at start of sampling and measurements.
NOT RECOMMENDED!
User skipped
SKIPPING THE AUTOMATIC TAPE ADVANCE AT
tape advance
THE BEGINNING OF SAMPLING WILL IMPACT
CORRECT OPERATION OF THE DEVICE. DATA
MAY NOT BE VALID.
Device stops sampling after receiving a command
System busy
via USB or serial that requires the device stop
operation.
Source apportionment disabled due to
S.A. disabled
measurement configuration. Check Settings to re-
enable source apportionment calculations
Tape is jammed. Inspect tape. Release and clamp
Tape jam
tape to attempt to clear the jam error.
Tape at end
Tape is at end. Replace tape.