Conditions for use GH3 Twin is a ceiling-mounted hoist that moves in a rail system. GH3 Twin is designed to be used with a variety of lifting hangers and lifting slings. GH3 Twin is designed to covers the requirements for heavy and special lifting or moving of a user.
The red strap for the emergency stop and the emergency lowering must be adjusted to the healthcare professionals reach, and must not be removed. • The GH3 must not be used where there is a risk of it being splashed with water. •...
Unpacking and Preparation Visual check of the GH3. If the GH3 is thought to be damaged upon reception, the GH3 must not be used before it has been checked and approved by Guldmann service team or by a Guldmnn authorized technician.
To do this, push the yellow reset button (see point 2.09). 1 .09 Transformer GH3 is equipped with batteries that require regular recharging. The transfor- mer for charging and the battery charging point must be connected by a qua- lified engineer or by Guldmann Service Team.
Installation of GH3 Cross hanger and Horizontal lifter Lifting hangers from other manufacturers Guldmann shall not be liable for faults or accidents that may occur as a result of using lifting hangers made by other manufacturers. If there is any doubt about the selection or use of a lifting hanger, please contact your supplier.
Page 10
Fig . 5 4. The GH3 Cross hanger is ready to use, when it is able to move freely in a horizontaI position (Fig. 6) Fig . 6 Installation of GH3 Horizontal lifter Use the oval hook to install the Horizontal lifter.
Plan the move. Avoid leaving the user in the lifting sling unattended. Working with the GH3 Twin The GH3 Twin runs easily in the rail system and does not have any special requirements for space or power in connection with moving. Attention can thus be fully focused on the users functional level and the healthcare professional technique.
2 .00 Description of functions Information panel on the GH3 bottom surface on leader hoist. Lamp indicator Information panel on the GH3 bottom surface on follower hoist. Lamp indicator 2 .01 Pictograms Emergency stop Emergency lowering function Reset emergency stop...
2 .02 Indicator lamps and audio signals Status Indicator Audio signals Possible GH3 lamps Functions – stand by All OK Green No charging Yellow, 3 x Beep after 60 sec after 15 sec Low battery Yellow Fault on hoist Yellow...
2 .03 Operation Hand control The GH3 Twin is switched on automatically when a button on the hand con- trol is pressed. The GH3 Twin is switched off automatically after approx. 8 minutes without activation. The USB in the hand control is only intended for connection of PDA/Netbook with Guldmann SIC software and may only be used by Guldmann service team or by a Guldmann authorized technician.
Page 15
Using tilt function on GH3 The GH3 has a tilt function. This function can only be used with a horizontal lifter. (See section 2.03 for positions of hand control buttons) 1. When the buttons 1 or 2 are pressed both hoists will lift or lower the horizontal lifter.
2 .04 Supplementary modules, GH3 Twin There are various supplementary modules for the GH3 Twin • CLM module (GH3 Twin with statistical function for management use) • Service module (GH3 Twin with Service module) • Scale module (GH3 Twin with integrated scale) Scale •...
Page 17
Next service Setup 2019.07.01 Setup Units Language Language 4-8 English Italianio Change Change Change Units Language 1-8 Language 5-8 Dansk Espanol Change Change Change Units Language 2-8 Language 6-8 Deutsch Francaise Change Change Change Language 3-8 Language 7-8 Svenska Portuguese BR Change Change...
Page 18
Menu structure, GH3 Twin with supplementary modules • Class III Scale • CLM module • Service module Scale Scale Info 1-2 Next service Version 01.00 2019.07.01 Start Next Show Scale Scale Info 2-2 Lifetime 80.3 Changes total lifts Tare Exit...
Page 19
Next service Setup 2019.07.01 Setup Language Language 4-8 Lifetime English Italianio total lifts Exit Change Change Language 1-8 Language 5-8 Lifetime Dansk Espanol heavy lifts Exit Change Change Language 2-8 Language 6-8 Last week Deutsch Francaise total lifts Exit Change Change Language 3-8 Language 7-8...
2.05 Configuration of supplementary modules, GH3 Twin Before the GH3 Twin is put into use, the hoist must be configured. Configu- ration covers language (Scale module/CLM module/Service module) and the unit for specification of weight (Scale module). Factory setting: Language:...
Shock effects The GH3+ with scale module includes high-sensitive sensors to register mass corresponding to the nominal recommended load. The sensors are highly sensitive and can be damaged by the effects of shock, for example pulling the GH3+ at extreme speed into an end stop.
2 .07 Scale module and Class III scale module The GH3 Twin with scale module (option) lets the healthcare professional determine the users weight. Scale Warning! If the scale is required to comply with the Non-automatic weighing instru- ments directive (EU Directive 2014/31/EU), a Class III scale must be used.
Page 23
Tare Exit 1. Press any key on the hand control to activate the hoist. Guldmann logo “G” appears in the display. Select “Menu” using the function key located immediately below the display. 2. Then select until the “Scale” menu appears in the display.
Page 24
Weighing with a scale Scale Scale 80.3 Tare Exit Tare Exit 1. Always reset the scale module before weighing. See section ”Resetting of the scale”. 2. Lift the lifting sling off the hanger. Place the user in the lifting sling before mounting it on the lifting hanger.
Page 25
Screen utilised when using the scale Start-up logo: Press ”menu” to select the function Menu Menu icon for scale: KG Scale Scale Menu icon for Class III scale: III Press the left navigation button ”Start” to enter the menu. Start Start Scale (in the scale menu): the scale starts up and resetting occurs Scale...
Page 26
(in other words, the load weighed is under 2 kg). Reset Exit Select ”Reset” and start over with setting the scale. An error has occured in the scale. The display shows an error code. Scale Contact Guldmann to help solve the problem. Reset Exit...
2 .08 Class III Scale (GH3+ with Class III Scale) Scale The GH3+ with scale module (option) provides the facility to determine the users weight. Start Description A Class III scale is a built-in digital scale that satisfies the requirements for...
2.09 Calibration/verification of Classe III Scale In order to maintain the medical approval the digital scale must be calibrated/ verified according to national regulatory requirements at an accredited testing institute. Software version and change counter are verified on the hand control’s dis- play.
(PDA/Net Book readouts requires a PDA/Net Book with Guldmann SIC/CLM software). The USB in the hand control is only intended for connection of PDA/Net Book and may only be used by Guldmann service team or by a Guldmann authori- zed technician. Number of lifts, total...
Page 30
1. Press any key on the GH3 hand control to activate the hoist. When the lifting module is activated, the display on the hand control is switched on and the Guldmann logo “G” appears. 2. Select “Menu” using the function key located below the display.
2 .11 Service module (GH3 Twin with service module) The GH3 Twin with Service module (option) saves all of the information about time and indication of next safety/service inspection. The GH3 Twin with Service module specifies the date of the next safety/service inspection.
Page 32
Pop-Up’s for Service module (supplementary module) There are two different Pop-Up’s (brief messages on the display) on the GH3 Twin with Service module. These pop-up’s notify the user of upcoming and exceeded dates for service inspections. Both Pop-Up’s appear immediately after the hand control has been switched on.
Constant pull: Emergency lowering is activated. Emergency stop If the GH3 Twin does not stop/react to the hand control when the GH3 Twin is in use, pull the red strap and the lifting/lowering functions (except emergency lowering) are deactivated. When the emergency stop is activated, the hoist will not function.
Page 34
Emergency lowering function, electric The emergency lowering function is identified as the essential performance. If the GH3 Twin fails, the electrical emergency lowering function is used to lower the user safely. The emergency lowering function is operated by a con- stant pull on the red strap that is used for the emergency stop.
The indicator lamp on the bottom of the hoist turns yellow if the charge status becomes low or if there is a complete interruption of the charging function. The GH3 Twin then has a limited number of lifts available before it must be re-charged.
The horizontal lifter may be used only for horizontal relocation of people lying on a level, horizontal surface. The user must be lying on a Guldmann sling or on another suitable sling. A level surface can be, for example, a bed, a stret- cher or the floor.
For long-therm storage, disconnect the battery plugs and the plug at the charging PC from the battery. 4 .03 How to prevent/avoid corrosion? When the GH3 Twin is mainly used in an corrosive environment, e.g. swim- ming pool, the hoist must be ordered with a special corrosion-preventive surface treatment.
Do not use the lifting sling if it is damaged or defective. Do not use the GH3 Twin if the lifting strap is damaged or defective. Contact your supplier and order a new lifting sling or a replacement of the lifting strap.
In connection with the purchase Guldmann may offer a service agreement for this inspection. The GH3 Twin with service module may only be serviced by the Guldmann Service Team or by a qualified service technician with access to the PDA/Net Book with Guldmann Service and Information consol.
6.00 Classification CE marking Medical equipment with respect to electrical shock, fire and mechanical hazards only. In accordance with UL 60601-1, CAN/CSA c.22.2 No. 601.1 Medical Device Class I in accordance with EU MDR Regulation Type B in accordance with UL/EN 60601-1 UK Responsible Person European Device Solutions Ltd .
Page 41
Transformer Class I IP20 Transformer Class II IP21 Examples of serial number label Lifting module GH3+ xxx xxx (xxxxx) max xxx kg / xxx lbs xxxxxx 33V AC, 2.5A, IP20 Duty cycle 2 min ON/18 min OFF xxxxx V. Guldmann A/S...
Page 42
230 V version Lifting hanger Hand control Part no. xxx xxx GH Lifting Hanger Edition max xxx kg / xxx lbs Date xxxx-xx-xx V. Guldmann A/S Graham Bells Vej 21-23A IP44 DK-8200 Aarhus N xxxxxx xxxxxx www.guldmann.com yyyy-mm-dd Barcode 128C / GS1-128...
8 .01 Module label, quick info A module label on the strap relief indicates which optional modules are built into the specific GH3 lifting module together with a clear indication of the Safe Working Load. SWL 350 kg/770 lbs Service...
Configuration of GH3 lifting module Basic configuration The GH3 lifting modules are configured in a large number of variants. The 6 first columns in the chart below describes the basic configuration: ceiling hoist family, specific type, Safe working load, number of lifting straps, number of lifting motors, number of horizontal drive motors.
Page 45
• • • Drive motors are not compatible with UL xx) Not compatible with UL Example: GH3+ 350 122 (xxxxx) 2 horizontal drive motors 2 lifting motors 1 lifting strap Safe Working Load SWL: 350 kg Ceiling hoist, type GH3...
Graham Bells Vej 21-23A, 8200 Aarhus N DENMARK In respect of Non-automatic weighing instrument designated GH3+ with variants of modules of load receptors and load cells. Accuracy class III, single-interval or multi range (dual) Maximum capacity, Max: From 200 kg to 700 kg...
Guidance and manufacturer’s declaration – electromagnetic emissions The GH3 is intended for use in the electromagnetic environment specified below. The customer or the user of the GH3 should assure that it is used in such an environment. Emissions test Compliance Electromagnetic environment –...
Page 51
Guidance and manufacturer’s declaration – electromagnetic immunity The GH3 is intended for use in the electromagnetic environment specified below. The customer or the user of the GH3 should assure that it is used in such an environment. IMMUNITY test IEC 60601...
Page 52
Guidance and manufacturer’s declaration – electromagnetic immunity The GH3 is intended for use in the electromagnetic environment specified below. The customer or the user of the GH3 should assure that it is used in such an environment. IMMUNITY test IEC 60601 TEST...
RF communications equipment and the GH3 The GH3 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the GH3 can help prevent electromagnetic interference by main-...
Guldmann Certified Technician. A Guldmann Certified Technician is a technician who has successfully completed Guldmann Service Training, and who holds a valid Service Training Certificate from Guldmann, and is in pos- session of a valid password to access Guldmann’s Service and Information Console (SIC).