4 Product description

4.1 Description of operation

The ENDOMAT
during operations in various disciplines of human and veterinary medicine. It is not possible to
combine disciplines of human and veterinary medicine.
The device automatically adapts to suit the type of operation being performed by providing the
optimum operating parameters when a tubing set for a specific discipline is attached, provided
the product is enabled for the area of application. The product can be configured in such a way
that only the areas of application that the user wants to use are displayed. Further disciplines
can be retrofitted, and extended setting options are available with the Advanced software
package.
In urological and arthroscopic applications, the irrigation pressure can also be increased briefly
with a BOOST function. In other applications, either the irrigation flow or the irrigation pressure
can be limited.
The product is operated and controlled via a touch screen. The current operating state can be
checked by displaying the set and actual values of the irrigation pressure or flow. If the set
value deviates continuously, an electronic safety circuit blocks the delivery or suction, and
acoustic signals sound. An electronic auto-check system tests the various system components
when the product is started and notifies the operator of any failures detected.
Suggested distension media
The ENDOMAT
1.5% and sorbitol 3.0%) and isotonic, electrolyte-containing media (e.g., saline solution 0.9%).
Automated fluid measurement system – EQUIMAT
For the automated fluid measurement system, we recommend the KARL STORZ EQUIMAT
device or the appropriate devices from other manufacturers.

4.2 Product overview

5
Fig. 1: ENDOMAT
Instructions for use • ENDOMAT
®
 SELECT is a roller pump that can be used for irrigation and suction of fluids
®
 SELECT can be used with hypotonic, electrolyte-free media (e.g., glycine
4
®
 SELECT – Front view
®
SELECT • 96326003EN-US • V 1.0
Product description
®
1
2
3
®
14