GENERAL
ELECTRIC
MEDICAL
SYSTEMS
AMX-il X-RAY UNIT
7-1576
SM A0654E
1.
Remove the piastic cover from the AMX Unit.
Unwrap the cabies tied to the horizontal
Telescoping Arm, but do not release or unstrap
the Telescoping Arm. Open the cassette drawer
and inventory the furnished parts refering to
Section 1-4 and copies of the Product Delivery
Instructions.
2.
Unpack the boxes containing Tube Unit,
Collimator, and accessory extension cylinder
assembly. inventory furnished parts refering
to Section 1-4 and Copies of the Product
Delivery Instructions.
Do not remove the two wooden shipping braces
fastened to the ends of the Tube Unit. They
will be removed after the Tube Unit is mounted
to the trunnion fork.
3.
Check the Tube Unit for an oil leak or a broken
insert.
Look for tell-tale oil slick on the housing
indicating an oil leak. In the event an oil slick
"is found at any of the "O" ring seals, tighten
the seal. Make sure the leakage is stopped.
Rotate and angulate the Tube Unit and listen
21979-P10
for any noise which would indicate a broken
ILLUSTRATION 4-11-2
insert. Using a flashlight examine the insert
,
-
.
through the Lexan Window.
ーー
tad:
COLLIMATOR INTERFACE
PLATE v,
NOTE:
The Tube Unit is shipped with the
collimator interface plate pre-
aligned and secured to the Tube
Unit faceplate.
Do not loosen any
of the (4) 10-32 unf Hex-HD screws
or move any of the excentrics which
secure the interface plate.
İf the
plate is shifted it must be realigned
following the procedure of Chapter
4 part 1, section 3-3.
411-9