AMX-11 X-RAY UNIT
7-1576
SM A0654E
7.
Carefully push the HV transformer into the cabinet with the HV receptacles facing the column.
ーー CAP
2
TEMPORARILY
TURN IN
#10 -32 SCREW TO
LIFT
OUT
İNSERT PUNCH TİP
.
|
'
AND TAP TO LOOSEN
REMOVE
AND DISCARD
O-RING
`
0
MA
12289-P10
ILLUSTRATION 3-9
8
Replace the HV Transformer retainer bracket shown in Illustration 3-8.
9.
Locate the HV transformer terminal cover and a plastic bag containing four 3" (76.2 mm) long spacer
studs.
Install the spacers on the four screws protruding out of the HV transformer.
Connect the six
wires to the HV transformer terminals as follows:
(Check that there is a 1" (25.4 mm) 1.D. plastic
grommet around the six wires before connecting.)
LEAD LABELS
CONNECT TO HV TRANSFORMER TERMINAL
P1
P1
P2
P2
G
G
XC
XC
XS
XS
MA
MA
10.
The terminal cover will be installed later, after tests are completed.
Set aside the cover and the four
8-32 unc x 1/2" machine screws furnished.