GE AMX-II Service Manual page 15

X-ray unit
Table of Contents

Advertisement

COLUMN
EXPOSURE
CONTROL
HAN DSWITCH
BATTERY
CHARGER
CORD &
PLUG
ILLUSTRATION 1-3. 23564P10
The BATTERY VOLTAGE
COMPENSATOR
SWITCH adjusts for variations in battery charge. When the
battery meter is in the HIGH region, the switch should be in the HIGH position. As the battery charge is
reduced to the LOW region, the switch should be in the LOW position. After battery charging is completed,
allow the following condition for at least 15 minutes before making exposures:
Keep the MAIN POWER
BREAKER
at ON. If the line cord is left connected, set the KEY SWITCH at CHARGE.
However, if the
line cord is disconnected, set the KEY SWITCH at ON.
The
Remote Exposure Contro! Handswitch (Illustration 1-3) has ROTOR and X-RAY buttons. Provided
that the red UNIT ON fight is glowing, depressing the rotor button will cause x-ray tube anode rotation
and filament preheating. There will be a 2.5 second delay for these functions to reach the operating point.
After that delay, the white READY
lamp will glow. Then the x-ray exposure will occur immediately after
the handswitch X-RAY button is depressed.
If the ROTOR
and X-RAY buttons are depressed simultaneously, the x-ray exposure will occur after the
2.5 second delay and as the READY lamp flashes on.
In order to obtain an exposure, the MAIN POWER
CIRCUIT BREAKER
must be ON, the X-RAY
GENERATOR
STANDBY SWITCH must be ON, and the charger cord disconnected from the power outlet.
Glow of the red X-RAY lamp indicates that x-rays are being emitted. When the READY
lamp is glowing,
the X-RAY
lamp will glow immediately after the handswitch X-RAY
button is depressed and it will remain
5