2.7

Automatic temperature compensation

The refractometer provides an automatic temperature compensation for the whole Brix
scale in the temperature range of 10–40 °C. This is permanently active. The
automatic temperature compensation corrects the measurement value in line with the
reference temperature of 20 °C.
 This means that regardless of the temperature at which a sample is measured
between 10–40 °C, the device displays the Brix value of the measured sample
for 20 °C.
The following sample illustrates this:
A sucrose solution with a value of 40 %Brix at 20 °C would have a value of
39,2 %Brix at 30 °C. If the sample is measured at 30 °C, the automatic temperature
compensation corrects the measured value and the refractometer displays 40 %Brix.
 The automatic temperature compensation is not used for the refractive index scale.

3 Unpacking and commissioning

3.1

Unpacking

N O T E
Risk of material damage during transport of the device
➢ Check device for transport damage after unpacking.
➢ Do not operate damaged devices.
When unpacking the device, check if any transport damage has occurred. Take
note of any loose parts, dents, scratches visible damage, etc.
Before disposal, inspect the packaging material for any loose functional parts and
safe if necessary.
Return the device in the original packaging with the completed return form; see
return form
(A.KRÜSS website).