2.7
Automatic temperature compensation
The refractometer provides an automatic temperature compensation for the whole Brix
scale in the temperature range of 10–40 °C. This is permanently active. The
automatic temperature compensation corrects the measurement value in line with the
reference temperature of 20 °C.
This means that regardless of the temperature at which a sample is measured
between 10–40 °C, the device displays the Brix value of the measured sample
for 20 °C.
The following sample illustrates this:
A sucrose solution with a value of 40 %Brix at 20 °C would have a value of
39,2 %Brix at 30 °C. If the sample is measured at 30 °C, the automatic temperature
compensation corrects the measured value and the refractometer displays 40 %Brix.
The automatic temperature compensation is not used for the refractive index scale.
3 Unpacking and commissioning
3.1
Unpacking
N O T E
Risk of material damage during transport of the device
➢ Check device for transport damage after unpacking.
➢ Do not operate damaged devices.
▪
When unpacking the device, check if any transport damage has occurred. Take
note of any loose parts, dents, scratches visible damage, etc.
▪
Before disposal, inspect the packaging material for any loose functional parts and
safe if necessary.
▪
Return the device in the original packaging with the completed return form; see
return form
(A.KRÜSS website).
Instruction Manual DR201-95
|
Version 3.1 © A.KRÜSS Optronic 2022
|
Page 11 of 30