5.4
Exchanging the battery
▪
It can be assumed that a 1.5 Volt battery will have a general service life of
approx. 5000 hours.
The voltage of the battery may be sufficient for normal use, but not for
measurements. In this case, the device will not take a measurement or provide
noticeably inaccurate readings.
Wide measurement discrepancies can be caused by a used battery.
In case of strong measurement deviations, the device should be calibrated with
distilled water after the battery has been exchanged,
and calibrating.
▪
If the error code "LLL" is shown on the display, the battery has to be replaced by a
new one.
Exchanging the battery
REQUIREMENT
▪
Switch off the device.
PROCEDURE
1. To exchange the battery, open the flap on the bottom of the device.
2. Remove the empty battery and insert a replacement in correct polarity.
3. Finally, close the flap.
RESULT
The battery has been replaced.
Adjust/calibrate the device with distilled water after exchanging the battery.
Instruction Manual DR201-95
|
Version 3.1 © A.KRÜSS Optronic 2022
see chapter 4.1 Adjusting
|