HEIDENHAIN PWM 8 Operating Instructions Manual page 48

Diagnostic set
Table of Contents

Advertisement

Analog Output signals (~ 11 µApp/1 Vpp)
Prepare the oscilloscope as described below:
Vertical deflection
Switch channels A and B to chop mode (CHOP).
Set the deflection coefficient (Sensitivity) of the
channels A and B to 0.5 V/DIV (11µApp), 0.2 V/DIV
(1Vpp).
Horizontal deflection
Set time coefficient (Time basis) to 0.5 ms/DIV.
Triggering
Trigger automatically (AUTO)
Trigger Channel A
Trigger positive edge
Calibration
Switch the input coupling switch (AC/DC/GND) of
the channels A and B to GND ( ç or 0)
Use the Y-position potentiometers of the channels
A and B to shift the electron rays congruently to the
center of the screen (fig.)
Switch the input coupling switch (AC/DC/GND) of
the channels A and B to DC
PWM 8 Settings analog signals
BNC
[°]
MODE
* Interface board 1 Vpp
.
BNC A
Ue1
BNC B
Ue2
25°
Measure amplitude
- 48 -
(A)*
(B)*
Possible oscilloscope display x - t or x - y