Performance Check Procedure
12. Vertical POSITION Range (all channels)
a. Set:
A SEC/DIV
0.1ms
VERTICAL MODE
CH 1 (ADD off)
CH 1 VOLTS/D1V
1 V
CH 2 INVERT
Off
SCOPE BW
Off
CH 1 and CH 2 Input COUPLING
AC
b.
Connect leveled sine-wave generator (SG 503) output to the CH 1 and
CH 2 inputs via a 50
Cl
BNC coaxial cable, a 50 H BNC termination,
and a BNC dual-input coupler.
c.
Position trace to center horizontal graticule line.
d.
Set leveled sine-wave generator output for two-division signal at
50 kHz.
e.
Set:
CH 1 VOLTS/DiV
0.1 V
CH 1 POSITION
Fully CW (clockwise)
f.
CHECK—that the bottom of the waveform is at least one division above
the center horizontal graticule tine.
g.
Set CH 1 POSITION fully CCW.
h.
CHECK—that the top of the waveform is at least one division below the
center horizontal graticule line.
I.
Set:
CH 1 POSITION
12 o'clock
VERTICAL MODE
CH 2 (CH 1 off)
CH 2 POSITION
Fully CW
j.
CHECK—that the bottom of the waveform is at least one division above
the center horizontal graticule line.
k.
Set CH 2 POSITION fully CCW.
l.
CHECK—that the top of the waveform is at least one division below the
center horizontal graticule line.
7-78