Rackmounting Instructions—7623A/R7623A Operators
WIDTH
Minimum space between the front rails of the rack is 17 - 5/8 inches. This space allows the slide-out
tracks to operate freely, permitting the R7623A to movein and out of the rack.
DEPTH
Total depth required to mount this instrument in an enclosed-cabinet rack is 24 inches. This allows
enough spacefor air circulation, signal connections, and the power cord.
The rear mounting brackets allow this instrument to be mounted in racks with 14-5/8 to 28-1/2
inches between the front and rear rails. The R7623A must be properly supported at the rear. Improper
support mayresult in instrument damage.
SLIDE-OUT TRACKS
The slide-out tracks provided with this instrument permit it to be extended out of the rack for main-
tenance and calibration without removing it from the rack. Be sure the power cord and signal cables are
long enough to allow operation in the extended position.
FRONT RAIL
OF
CABINET RACK
Figure 3-2.
Vertical mounting position of the left stationary section and the securing hole location. These dimensions
also apply to theright frontrail.