4. Maintenance and repair
4.1.
Problem
My device switches itself off.
My device is blocked, the force no
longer varies.
My device doesn't display the
measurement screen but an
information page with Offset in red.
My device does not display the
measurement screen but an
information page with Overload
greater than 10
My device does not display the
measurement screen but an
information page with Recalibration
date in orange
Main faults and solutions
Solution
1. Check that the battery percentage is not at 0%.
2. The Centor EASY II switches off automatically after 15
Switch off and restart your instrument using the on/off
button.
If the offset value is greater than 20%, the device cannot
measure. Switch off your device and remove the load
applied to the sensor. Then switch on your device.
If no load is applied to the sensor, then the sensor is
deformed. In this case, return the device to us for
replacement.
You can reset the overload counter to 0 from the Admin
menu. To access it, press and hold the "U" key while in
the Main menu or Autotest. The Admin menu appears.
Go to the Overload line and press the right arrow.
Confirm that you wish to reset overloads to 0. Your
device is now ready for operation again.
You can continue measuring. If you want to change the
date of the next calibration, go to the Admin menu. To
access it, press and hold the "U" key while in the Main
menu or Autotest. The Admin menu appears. Move to
the Next Cal line and press the right arrow. Use the up
and down arrows to change the date. Your device is now
ready to use again.
Charge the device.
minutes if no key is pressed.
- 21 -
Instruction manual
CENTOR EASY II