To change the type of value to be saved, move to the "Peak" line, then press the right arrow. A message
appears, asking if you wish to delete all values in memory. Warning: changing the type of value to be
saved systematically deletes all values in memory.
3.4.2. SET POINTS menu
If you wish to clear the memory without changing direction, press and hold the left arrow.
threshold is exceeded and/or emits a sound.
How it works:
•
If Value 1 and Value 2 are 0, the threshold function is deactivated.
•
If Value 1 has a value and Value 2 is 0. Only one threshold is activated. In this case, if the
maximum of the chosen direction exceeds the value in Value 1, the force gauge beeps and/or
the digits turn orange.
If Value 1 and Value 2 have a value, then there are 2 thresholds set. In this case, if the maximum
•
force exceeds Value 1, the Centor Easy II emits a beep and/or the values turn green. If the
maximum force exceeds Value 2, the values change to orange.
Example of use:
1. I've got a 500N force gauge, and I want to shoot a sample. If it breaks above 100N, my parts
are good (displayed in green). The settings to be applied are as follows:
Mode: Traction
Value 1: 100N
Value 2: 500N
Color: Yes
Beep: No
2. I have a 100N force gauge and I want to compress a sample. If the force exceeds 50N, my parts
are bad (orange display and beep). The settings to be applied are as follows:
This menu allows you to set thresholds on the force
gauge. A sound can be emitted when thresholds are
exceeded, and the color of the measurement display
can be changed according to thresholds.
Mode: selects whether the threshold is in tension or
compression, clockwise (CW) or counterclockwise
(CCW).
Value 1 and Value 2: threshold values in the unit of
measurement.
Color, BIP: changes the color of measurements if the
- 17 -
Instruction manual
CENTOR EASY II
Need help?
Do you have a question about the CENTOR EASY II and is the answer not in the manual?
Questions and answers