What to do if...
If alarm occurs:
Mute alarm (if necessary)
Detect the cause and proceed to the appropriate action
Release alarm
Resume feeding
Empty battery
Connect pump to
mains supply
immediately.
Door open alarm
Close pump door.
Set installation alarm
Check set position
above and below
pump mechanism
and insert it correctly.
Fresenius Kabi Limited,
Cestrian Court, Eastgate Way,
Manor Park, Runcorn,
Cheshire WA7 1NT
tel: 01928 533533 fax: 01928 533520
email: scientific.affairsUK@fresenius-kabi.com
www.fresenius-kabi.com/gb
Date of preparation: March 2018 EN01046
24535 Fresenius Amika EN01046.indd 1-2
Start Reminder
Proceed to next step
or switch pump off.
Empty bag/ air in
Downstream
set alarm
occlusion alarm
Change container
Check that pump
and set or check for
set / feeding tube is
air bubbles.
clear after the
pump, at patient
side.
Upstream
Target volume
occlusion alarm
almost reached
message
Check that pump
set is clear between
Continue feeding
the container and
or end feeding.
the pump.
Target volume
Pump error alarm
reached alarm
Note the technical
End feeding.
Error code (Err xyz)
and contact your
biomedical
department.
Amika
®
is a registered trademark of Fresenius Kabi AG.
Fresenius Kabi Ltd is an authorised user © Fresenius Kabi Ltd.
Amika
®
ENTERAL NUTRITION PUMP
PICTORIAL REFERENCE GUIDE
08/03/2018 08:53