Section Vili
THEORY OF OPERATION
Model 3455A
8-30. 1 V de Reference Offset Error Measurement.
Since
the 1 V dc full scale reference was derived by dividing the
internal
reference by the precision 'ten-to-one divider, a
separate offset Error measurement is made to include any
offsets present in the ten-to-one
divider and associated
@
circuitry.
Figure 8-12 illustrates the circuit configuration
3455-8-4531
3455-8-4531
Figure 8-10.
10 V de Gain Error Measurement.