HP 3455A Operating And Service Manual page 18

Digital voltmeter
Hide thumbs Also See for 3455A:
Table of Contents

Advertisement

Section I
1-8
Model 3455A
Table 1-3. Recommended Test Equipment.
Instrument
Critical Specification
DC Voltage Standard
AC Calibrator
Test Oscillator
Resistance Decade
OC Nult Voltmeter
Reference Divider
DC Transfer Standard
Electronic Counter
Resistance Standard
Bus System Analyzer
Calculator
Oscilloscope
Digital Voltmeter
Resistors
Signature Analyzer
10 mV to 1000 V
+ .005%
Voltage:
Accuracy:
Frequency:
20 Hz to 100 kHz
Output Level:
100 mV to 1000 V
Accuracy:
+ .1%
Voltage Stability {6 mos.)
+ .02%
Frequency:
to 250 kHz
Output:
3 V rms into 50 2
Frequency Response + .25%
Resistance:
1009 to 10 M2
Accuracy:
+ .004%
Voltage Range:
1 nV to 10 V
Division Ratio Accuracy + .001%
Output Voltage Range
-
1 V to1kV
Output Voltages:
1 V, 1.018 V,
1.019 V,10V
Accuracy:
+ 5 ppm
Stability:
+ .001%
(30 days)
50 Hz to 6O Hz
Resistance:
1 kQ
Accuracy:
+ .0005%
Resistance:
100 K
Accuracy:
+ .002%
HP-|B Control Capabitity
HP-IB Control Capability must
serve
as printer for 3455A Output data.
Bandwidth:
DC to 10 MHz
Sweep Time: 0.1 ys to 1 sec/div
Sensitivity:
1 V/div
Voltage Range:
10 mV te 1000 V
Resolution:
10 2V
Resistances:
1k
+ 10%
10 kD + 0.1%
1 MQ + 0.1%
Recommended Madet
Use
Systron Donner
PAT
Model M107
-hp- Model 745A
PAT
AC Calibrator
-hp- Model 746A
High Voltage Amplifier
-hp- Model 652A
P
Test Oscillator
Gen Rad Model
PAT
GR 1433-Z Decade
Resistor
-hp- Model 419A
| PAT
Fluke Model 750A
PA
Reference Divider
Fluke Model 731A
PA
DC Transfer Standard
-hp- Model 5300A/5302A
P
Measuring System
Guildline Modet
A
9330/1
K or 9330A/1
K
Guildline Model 9330/100K
-hp- Model 59401A
T
Bus System Analyzer
-hp- Model 9825A
OT
-hp- Model
180C/D
Oscillo-
| T
scope
with
1801A
and
1821A plug-in units.
-hp- Model 3490A
-hp- Part No,
0684-1021
0698-4157
0698-6369
-hp- Model 5004A
erformance Checks
djustments
Troubleshooting
T=
O = Operators Check

Advertisement

Table of Contents
loading

Table of Contents