OPERATIONAL VERIFICATION TEST CARD (Cont'd).
OHMS ACCURACY
TEST
Decade
Ohms
Resistor
Function
Test Limits*
"Record 24-hour or 90-day test Jimits from table designated in test
procedure.
Derive 6-month test limits from specifications listed in
Table 1-1.
OC VOLTMETER
INPUT RESISTANCE
TEST
3455A
Test
Range
Reading
Test Limits
10V
——v__
| 9.99900 V to 10.0000 v
100 V
——V_
| 9.0900 V to 9.0917 V
&