M A N U A L A D J U S T M E N T P R O C E D U R E H M 1 5 0 8 - 2
3

Trace Rotation Check

Select "Trace Rot.".
Set baseline with POSITION 1 to the horizontal centre line of the graticule.
Turn INTENS knob to adjust baseline exactly parallel to the horizontal centre line of the graticule.
4
R9057
Readout Operating Point
Locate and identify R9057 on YF-Board.
Connect a 25 mV
, 10 Hz square wave signal via 50 cable and 50 through terminator to input CH 1.
pp
Set input attenuator CH 1 to 1 mV/div.
Set input coupling CH 1 to AC.
Set time base to 2 ms/div.
Adjust R9057 for minimal fl ickering readout display.
56
Subject to change without notice
C9001
C9037
R9034
R9126
R9114
C9045
R9118
R9057
C9009
R9149