Performance, Operating And Storage Conditions; Electrical Data; Materials Exposed To Vacuum - Edwards AIM-S-NW25 Instruction Manual

Active inverted magnetron gauge
Hide thumbs Also See for AIM-S-NW25:
Table of Contents

Advertisement

NRA737880 Issue E
2.2

Performance, Operating and Storage Conditions

Ambient temperature
Operation
Storage
Ambient humidity (operation)
Maximum operating altitude
Maximum internal pressure
Pressure measurement range
Pollution category
2.3

Electrical Data

Electrical supply
Voltage
Max voltage ripple
Max source resistance
Maximum power consumption
Electrical connector
Pressure output signal
Range
Error range
dDimensions(mm).c
Impedance
Min load impedance
Max current source
Gauge enable
Control sense
Active level
Control impedance
Error output
External load rating
Back EMF suppression diode*
Min. surge rating
Min. reverse voltage rating
Gauge identification resistance
* Required when you see an external d.c. relay connected to the error output.
2.4

Materials Exposed to Vacuum

Stainless steel (AISI 304 and 306)
Fluoroelastomer
Soda lime glass.
Page 4
º
5 to 60
C
º
0 to 70
C
10 to 90% (non-condensing)
2000m
6 bar absolute (5 bar gauge)
-8
-2
1 x 10
to 1 x 10
mbar
EN61010 Category 2
+13.5 to +36 V d.c.
1 V peak to peak
50 W
3.5 W
FCC68/RJ45 type, 8-way
2
output
10 V d.c.
output < 2 V d.c. or output > 10 V
Ω
0.1
Ω
10 k
μ
1
A
Active low
< 1.3 V
12 kW pull-up to positive supply
40 V d.c., 100 mA max
1 A
100 V
100 kW ± 2%
Edwards and the Edwards logo are trademarks of Edwards Limited.
© Edwards Limited 2020. All rights reserved.

Advertisement

Table of Contents
loading

This manual is also suitable for:

Nra737000