NRA737880 Issue E
The connection to pin 6 is optional. Make the connection to pin 6 if you want to connect the error signal to a d.c.
relay: you must connect a suppression diode between pins 1 and 6 to protect the AIM Gauge from transient voltages
generated when the d.c. relay is switched off.
Connect a switch between pins 2 and 7 to enable and disable the gauge (refer to
Section
4.2). If you want to measure
the gauge identification signal (which identifies the AIM Gauge), measure the resistance between pins 4 and 5.