Warning: For safe operation, the humidi昀椀er must always be positioned below the breathing circuit connection
at the mask. The humidi昀椀er must be level for proper operation.
Note: Refer to the humidi昀椀er's instructions for complete setup information.
Using the SD Card
The DreamStation system comes with an SD card inserted in the SD card slot on the side of the device to record
information for the home care provider. Your home care provider may ask you to periodically remove the SD card
and send it to them for evaluation.
Updating Software Using the SD card
To check which version of software is currently on your device, navigate to My Provider and select Device Info.
You can update the device software using the SD card. The software update must be done when the therapy is off.
Insert an SD card with the new software version into the device. A pop-up screen appears asking "Would you
1.
like to upgrade software?"
2.
Turn the control dial to select Yes and then press the control dial to start the upgrade. The busy icon appears
while the upgrade is in progress. Do not remove power from the device.
If the software update is successful, the Change Accepted icon appears on the screen. Removed the SD card
3.
from the device to restart the device and use the new software.
4.
If an SD card error is detected, the Change Rejected icon appears . Remove the SD card and reinsert. If the
alert continues to occur, contact Philips Respironics at 1-800-345-6443 or 1-724-387-4000 for a new SD card.
Using the DreamStation Link Module
The Link Module is able to receive oximetry data and transfer it to the therapy device for home use or in a laboratory
setting. For use in a laboratory setting, the Link Module also includes an RS-232 (or "DB9") port to allow remote
control of the DreamStation Sleep Therapy Device by a personal computer.
Note: Please consult the instructions that accompany the Link Module for installation and removal.
Note: There are no SpO
Note: Oximetry data is not displayed.
Dispose of the module following the same disposal instructions for your therapy device.
Warnings:
• If you notice any unexplained changes in the performance of this device, if it has been dropped or mishandled,
if water is spilled into the enclosure, or if the enclosure is broken, discontinue use. Contact your home care
provider.
• Repairs and adjustments must be performed by Philips Respironics-authorized service personnel only.
Unauthorized service could cause injury, invalidate the warranty, or result in costly damage.
• Do not use any accessories, detachable parts, and materials not recommended by Philips Respironics.
Incompatible parts or accessories can result in degraded performance.
Adding Supplemental Oxygen
Oxygen can be added to the patient circuit. Please note the warnings listed below when using oxygen with the device.
Warnings:
• When using oxygen with this system, the oxygen supply must comply with local regulations for medical oxygen.
• Oxygen supports combustion. Oxygen should not be used while smoking or in the presence of an open 昀氀ame.
• If supplemental oxygen is added at the exit of the 昀氀ow generator or humidi昀椀er, a Philips Respironics Pressure Valve
must be placed in-line with the patient circuit between the device and the oxygen source. The pressure valve helps
prevent the back昀氀ow of oxygen from the patient circuit into the device when the unit is off. Failure to use the
pressure valve could result in a 昀椀re hazard.
• When adding oxygen at the mask end of the hose, a Philips Respironics Pressure Valve is not required for oxygen
昀氀ow rates of ≤4 liters per minute. However, the reusable and disposable 昀椀lters must be in place on the 昀氀ow
generator. Failure to install both the reusable and disposable 昀椀lters could result in a 昀椀re hazard.
Note: Refer to the pressure valve's instructions for complete setup information.
• When using oxygen with this system, turn the device on before turning on the oxygen. Turn the oxygen off
before turning the device off. This will prevent oxygen accumulation in the device.
• Do not connect the device to an unregulated or high pressure oxygen source.