Philips Respironics DreamStation CPAP User Manual page 20

Table of Contents

Advertisement

Alert
Icon
SD Card
Removed.
Oximetry: Good
Study
(icon only)
SD Card Error:
Remove and
Reinsert
SD Card Full.
Patient Message
(Refer to section)
Change Rejected
Humidi昀椀cation
Error. Contact
support if the
problem persists.
18
Type
Description
Noti昀椀cation
Indicates SD card has
or Alert 2
been removed from
therapy device and not
reinserted before the
start of the current
therapy session.
Noti昀椀cation Noti昀椀es that user has
a achieved at least 4
hours of therapy and
oximetry use. Appears
at the end of therapy.
Noti昀椀cation SD card error detected Device cannot read
Noti昀椀cation SD card is full.
Noti昀椀cation Message from your
Provider.
Alert 1
A prescription or
settings change was
rejected.
Humidi昀椀er error (only
Status
when humidi昀椀er is
present)
Possible Cause
SD card was not
reinserted into device.
n/a
the SD card. A
problem may exist
with the SD card or
it was ejected during
a writing activity,
or it was inserted
incorrectly.
SD card is full.
n/a
Change missing or
incorrect.
Humidi昀椀er heater
plate error or
humidi昀椀er not
properly connected to
therapy device
Action
Reinsert SD card, or
click to clear alert.
Press Control Dial to
acknowledge and clear
the message.
Remove SD card
and reinsert. If alert
continues to occur,
replace with another
card or contact your
provider.
Remove SD card and
replace with a new
card, or contact your
provider for a new SD
card.
Press Control Dial to
acknowledge and clear
the message.
Contact your provider.
Turn off device and
disconnect from
power. Detach the
humidi昀椀er, visually
check that electrical
contacts are clear,
then reconnect
humidi昀椀er and
power cord. If alert
continues, contact
your provider.