AMPROBE
Fluke Corporation
Product : HD160C
AUTO CALIBRATION
Equipment Required :
Multifunction Calibrator - Fluke 5500A or equivalent
Standard Capacitors 10 nF,100 nF,1 uF,10 uF,100 uF
Caution:
and will be used in parts of the following procedure. This procedure should
Adjustments :
Step 1 : Set the Function Switch to OFF position. Open the rear case of the meter.
Move the jumper, mounted on PC board, from 'Normal' (J2) to 'Calibration' (J1) position.
Reattach the battery.
Step 2 : DCV Adjustment - Set the function switch to DCV range.
2-1. Zero Adjustment on DC 1000 mV range
Apply short bar to V - Ω and COM jacks and let reading settle
Press HOLD button ( a beep sound is audible ) to save setting.
LCD will indicate 000.0 mV ±1 digit.
Dangerous voltages and currents are available from calibrators
be done by a qualified technician
+
Calibration
Normal operation
1 of 4

Advertisement

loading
Need help?

Need help?

Do you have a question about the AMPROBE HD160C and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Fluke AMPROBE HD160C

  • Page 1 Fluke Corporation Product : HD160C AUTO CALIBRATION Equipment Required : Multifunction Calibrator - Fluke 5500A or equivalent Standard Capacitors 10 nF,100 nF,1 uF,10 uF,100 uF Caution: Dangerous voltages and currents are available from calibrators and will be used in parts of the following procedure. This procedure should...
  • Page 2 AMPROBE Fluke Corporation Product : HD160C 2-2. DC 1000 mV Adjustment Apply + 900 mVDC to V - Ω and COM jacks and let reading settle Press HOLD button ( a beep sound is audible ) to save setting. LCD will indicate 900.0 mV ±1 digit.
  • Page 3 AMPROBE Fluke Corporation Product : HD160C Set the function switch to uA range Apply + 90 uA to A and COM jacks and let reading settle Press HOLD button ( a beep sound is audible ) to save setting. LCD will indicate 90.00 uA ±1 digit.
  • Page 4 AMPROBE Fluke Corporation Product : HD160C 5-3. ℃ Adjustment Using the ‘RANGE’ button,clear the LCD of MAX or MIN symbols. Connect the XR - TA adapter to the V / Ω and COM jacks ( + to V / Ω) Apply 0℃( from Calibrator ) via K type connector and let reading settle...