Page 1
Fluke Corporation Product : HD160C AUTO CALIBRATION Equipment Required : Multifunction Calibrator - Fluke 5500A or equivalent Standard Capacitors 10 nF,100 nF,1 uF,10 uF,100 uF Caution: Dangerous voltages and currents are available from calibrators and will be used in parts of the following procedure. This procedure should...
Page 2
AMPROBE Fluke Corporation Product : HD160C 2-2. DC 1000 mV Adjustment Apply + 900 mVDC to V - Ω and COM jacks and let reading settle Press HOLD button ( a beep sound is audible ) to save setting. LCD will indicate 900.0 mV ±1 digit.
Page 3
AMPROBE Fluke Corporation Product : HD160C Set the function switch to uA range Apply + 90 uA to A and COM jacks and let reading settle Press HOLD button ( a beep sound is audible ) to save setting. LCD will indicate 90.00 uA ±1 digit.
Page 4
AMPROBE Fluke Corporation Product : HD160C 5-3. ℃ Adjustment Using the ‘RANGE’ button,clear the LCD of MAX or MIN symbols. Connect the XR - TA adapter to the V / Ω and COM jacks ( + to V / Ω) Apply 0℃( from Calibrator ) via K type connector and let reading settle...