Advanced Configuration
SETTINGS
Settings has 4 categories (Range Adjust, HART, Level Variables and Signal Properties)
Write to Unit is used to save any changes to the sensor. Any changes made are not saved into EchoBeam
until Write to Unit is pressed.
|
56
LEVEL VARIABLES
Multiwave
o Used when there are multiple return echoes displayed in the
EchoCurve.
o Selections are YES or NO with NO as the default.
Section Eight
Below is one valid echo that
is above the envelop curve
and is wide with the actual
level being at the highest
point (7.028 ft).
With Multiwave ON, the
echo is separated, and the
First Echo becomes the
actual level distance (6.333
ft).
MN310255 Rev 10
®