Appendix
Register Mapping:
0x0000
Distance measurement in centimeters (read only)
0x0001
Distance measurement in millimeters (read only)
0x0002
Level measurement in centimeters (read-only)
0x0003
Level measurement in millimeters (read-only)
0x0004
Measurement status (read-only);
= Normal
0
= Comm Err - Display not connected or broken
1
= EEProm Err - write failure, EEPROM is broken
2
= PLL Err - antennae circuit is broken
3
= SYSBUSY - Slow response due to large data request or low supply voltage
7
= LOWPOWER - Low Voltage - check power supply
8
= NOSIGNALERR - No valid echo signal
14
0x0005
Range Value in centimeters (read and write);
0x0006
Blanking distance (dead zone) in centimeters (read and write);
0x0007
Low-level (Sensor Height) adjustment in centimeters (read and write);
0x0008
High-level (Fill-Height) adjustment in centimeters (read and write);
0x0009
Signal Strength (read-only)
= Min
0
= Max
150
0x000A
Distance measurement in millimeters (4-byte Integer, read-only, high 16 bits)
0x000B
Distance measurement in millimeters (4-byte integer, read-only, low 16 bits)
0x000C
Distance measurement in meters (4-byte float, read-only, high 16 bits)
0x000D
Distance measurement in meters (4-byte float, read-only, low 16 bits)
0x000E
Level measurement in millimeters (4-byte integer, read-only, high 16 bits)
0x000F
Level measurement in millimeters (4-byte integer, read-only, low 16 bits)
0x0010
Level measurement in meters (4-byte float, read-only, high 16 bits)
0x0011
Level measurement in meters (4-byte float, read-only, low 16 bits)
0x0012
Temperature (4-byte float, read-only, upper 16 bits)
0x0013
Temperature (4-byte float, read-only, lower 16 bits)
|