Parameters

Once entering in to the parameter—input mode, set the parameters as the order shown below.
After finishing set all the parameters, it can come into the normal operation.
Parameter input Mode
Fahrenheit Temperature
Display

Calibration of Iron Temperature

Temperature should be recalibrated after changing the iron, replacing the heating element and the iron
tip. The methods of recalibrating iron temperature as followings with a tip thermometer. This method
is greater accuracy.
Calibration with a Tip
Thermometer.
1. Turn off the power switch. Press and hold the "up" and
"down" buttons simultaneously, then turn on the power
switch.
2. Continue holding down the "up" and "down" buttons
until the display shows F (Fahrenheit).
3. When the window displays F, the station is in
parameter—input mode.
After selecting F temperature display, the heater error
temperature will be displayed and flashing at the 100's
digit.
See Heater Error. Set the heater — error temperature-
tolerance in the same manner as described in "setting the
temperature" (steps 2—4). Be sure to use a value in the
allowable range.
1. Set the temperature at 750°F.
2. Wait till the temperature stabilizes and remove out the
CAL pot plug.
3. When the temperature stabilizes, use a (-)-type or (+)-
type screwdriver to adjust the screw (marked CAL at the
station) until the tip thermometer indicates a temperature
of 750°F. Turn the screw clockwise to increase the
temperature and counterclockwise to reduce the
temperature.
After calibration, insert the CAL pot plug again.
* We recommend the 191/192 thermometer for measuring
the tip temperature.