GENERAL CLEANING

After each procedure, the table should be carefully cleaned
The disinfectant cleaners and cleaning steps on this page apply to all the parts of the
table except the radiation shield. (see the following page for radiation shield cleaning)
• Clean tabletop, pads, accessories, and painted surfaces with a clean cloth dampened
with an approved cleaner (see list below).
Do not use abrasives, solvents, sprays, or corrosive agents.
APPROVED AND TESTED DISINFECTANT CLEANERS
• Sodium hypochlorite (generic household bleach) in solution of 5.25% sodium hypochlorite
diluted between 1:10 and 1:100 with water.
• Alcohol (generic)
• Envirocide ® Disinfectant and Cleaner
CLEANING STEPS
1. Move the tabletop into the horizontal position.
2. Lower the tabletop to the lowest position.
3. Power cord, hand control and foot control cords must be plugged in to
protect the sockets.
4. Remove all pads and accessories.
5. Wipe off any access fluids with a water dampened cloth or sponge.
6. Clean the accessories, tabletop, side rails, frame and pads using an
approved cleaner.
7. Thoroughly rinse tabletop pad with water.
8. Gently rub with a soft, clean cloth to dry.
If noise or uneven motion occurs when the table top is moving up or down,
extend and clean the lifting tube with a soft clean cloth.
L100-3065 REV C
Aspect 100UC Plus
Avoid cleaning or getting cleaning solutions on elevating
www.imagediagnostics.com
CAUTION
tube if possible.