Quick Start Guide
2.3.2  Sensor orientation
Install all Rosemount pH/ORP sensors within 80 degrees of vertical,
as shown in
sensor does not stay at the sensor tip and disrupt the measurement.
Figure 2-1: Sensor orientation
Install the sensor tip down within 80 degrees of vertical.
2.3.3  Install insertion style sensors
Emerson offers two primary mounting styles for pH/ORP sensors:
insertion and retractable.
Insertion style sensors have NPT pipe threads that you can connect
directly into the process or into a mounting adapter.
Figure 2-2: Example of a Rosemount insertion style sensor
A. Forward facing ¾-in. NPT are present on some Rosemount pH/ORP
sensor models.
B. Forward facing 1-in. NPT
C. Wrench flats
D. Rear facing 1-in. NPT
You can connect insertion style sensors to the process using the
forward facing NPT threads or you can use the rear facing NPT
threads for immersion mounting in a pond or basin.
6
Figure
2-1. This ensures that the air bubble within the
May 2023
Rosemount pH/ORP Sensors