Reference Guide Correct Usage of EASYSPRAY and DUPLOSPRAY Pressure Regulator...
Page 2
EASYSPRAY and DUPLOSPRAY EASYSPRAY Pressure Regulator Operation (open surgery) Insert 9V battery into the Connect Spray Set filters EASYSPRAY Pressure to EASYSPRAY device. Regulator device Connect the blue filter to the blue female luer connector and the clear filter to the...
Page 3
MIS Pressure Regulator - Instructions for correct usage CAUTION BOX Any application of pressurised gas may be associated with a potential risk of air or gas embolism, tissue rupture or air or gas entrapment with compression, which may be life threatening if the product is sprayed incorrectly Precautions For TISSEEL Fibrin Sealant and ARTISS Solutions for Sealant...
Page 4
Fibrin Sealant Surgery Spray set to be used Applicator tips to be used TISSEEL/ARTISS Spray Set n.a. Open surgery TISSEEL/ARTISS Spray Set 10 Pack n.a. DUPLOSPRAY MIS Applicator 20cm DUPLOSPRAY MIS Applicator TISSEEL 30cm Laparoscopic/ minimally n.a. invasive procedures DUPLOSPRAY MIS Applicator 40cm Replaceable Tip TISSEEL/ARTISS Spray Set...
Page 5
Recommended Pressure and Distance | Overview Recommended distance Pressure regulator to be used Recommended spray pressure/flow rate from target tissue EASYSPRAY 10-15cm 1.5-2.0 bar (21.5-28.5 psi). EASYSPRAY DUPLOSPRAY MIS Regulator DUPLOSPRAY MIS Regulator NIST B11 DUPLOSPRAY MIS Regulator DUPLOSPRAY MIS Regulator NIST B11 1.0-2.0 L/min...