THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT.
PRODUCT CODE: 2M8153, 2M8153K
.
Prior to using this pump, read this manual carefully to
fully understand the pump's functionality and to ensure
safe and proper operation.
Printed from the Baxter Intranet.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Colleague 3 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for baxter Colleague 3

  • Page 1 Prior to using this pump, read this manual carefully to fully understand the pump’s functionality and to ensure safe and proper operation. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT.
  • Page 2 There are risks associated with using anything other than the recommended sets with this device. Sets designated for use with this device are identified in Chapter 3, Administration Sets. Baxter’s warranty on this device will be null and void and Baxter will assume no responsibility for incidents which may occur if the product is not used in accordance with product labelling.
  • Page 3: Table Of Contents

    In EEC Countries ..................In Latin America ..................In the United States..................Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 4 Checking Battery Charge Level ................4-5 Using the Configuration/Service Function ............4-6 Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 5 Personality Feature Sets..................5-25 Overview....................5-25 Selecting a Pump Personality Feature Set ..........5-26 Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 6 Battery Operation Overview ................. 7-3 Battery Charging ..................Battery Disposal ..................Storage ........................7-4 Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 7 ..........Index-1 Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 8: Material Specifications

    THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 1. Pump housing components may be made of PC/ABS or Acetal. Some pumps may contain PC/ABS and Acetal components. COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 9: Meaning Of The Ce Mark Symbol

    Deerfield, IL 60015 USA Made in Singapore Authorised Baxter S.A. Representative: B-7860 Lessines, Belgium Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 10 Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT.
  • Page 11: Chapter 1 Introduction

    Introduction Chapter 1 Overview The Colleague 3 Volumetric Infusion Pump is designed to meet the fluid delivery needs of today’s evolving healthcare environment. The Colleague 3 Pump can be utilized for continuous or intermittent delivery through clinically acceptable routes of administration such as, intravenous (IV), intra-arterial (IA), subcutaneous, epidural, or irrigation of fluid spaces applications.
  • Page 12 Overview The Colleague 3 Pump feature spectrum includes: Micro and Macro Rate Range ■ Basic Delivery Programming ■ Piggybacking Secondary Medications ■ Special Programming Functions for Dosing ■ The following pump features help provide safe, efficacious, and reliable therapy: Automatic tube loading with misloading detection ■...
  • Page 13 Introduction Overview Three independent Pump Channels for infusion. U.S.: Uses Baxter standard administration sets equipped with keyed On/Off Clamps. EEC COUNTRIES: Uses Baxter Green Label C96XX administration sets equipped with keyed On/Off Clamps. See “Recommended Administration Sets,” 3-3. Automatic Tube Loading...
  • Page 14 Configurable Pump Personality Feature Sets Battery Charge Level Indicator Nurse Call Communications Port Diagnostics Functions Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 15: Safety Summary

    ■ manual to fully understand the functionality and to ensure safe and proper operation. Although the Colleague 3 Pump has been designed and ■ manufactured to exacting specifications, it is not intended to replace trained personnel in the supervision of IV infusions.
  • Page 16 8060517CT manufactured June 1998 9081234CT manufactured August 1999 10025678CT manufactured February 2000 Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 17: Warnings

    When disposing of this device or the administration sets designed for use with ! WARNING ! this device, adhere to local regulations and guidelines. This pump should be used only with Baxter accessories specified in “Using the ! WARNING ! Optional Accessories,” 5-24. There are risks associated with using anything other than the recommended accessories with this pump.
  • Page 18: Cautions

    Safety Summary The Colleague 3 Pump is intended for use in delivering multiple infusions to a ! WARNING ! single patient. Never use the pump to deliver infusions to more than one patient simultaneously. If the pump has been dropped or appears to be damaged, it should be taken out ! WARNING ! of service and inspected by qualified service personnel only.
  • Page 19: Notes

    Certain combinations of error codes can result in default to the English language. If this should occur, dis- continue using the pump and contact your authorized Baxter Ser- vice Center for assistance. Note: U.S.
  • Page 20 Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT.
  • Page 21: Chapter 2 Pump Description

    Overview This chapter describes the operator controls and indicators on the pump. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 22: Main Body

    PUMP MODULE A PUMP MODULE B PUMP MODULE C Figure 2-1 Front View Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 23 Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 24 (A, B, C). Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 25: Main Display

    These keys are used to select programming fields or to perform actions. Arrow Keys Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 26: Pump Modules

    If the pump is running on battery power with no alarm or alert, the pump module display is blank to conserve energy. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 27 These arrows indicate the direction of fluid flow. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 28 This LED is lit when the associated pump channel has been selected. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 29: Rear Panel

    Communication port An RS232/RS423 interface enables optional nurse call and communication functions. The pump can be monitored by a computer. For information on computer monitoring, refer to the Colleague Pump Computer Monitoring Guide, available from your Baxter customer representative. Fuse holders The pump’s fuses are located inside.
  • Page 30: Display Reference Guide

    Soft Key Identifier Prompt Soft Keys Figure 2-5 Main Display Information Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 2-10 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 31: Programming Screens

    Select soft key. Figure 2-7 Programming Modes Menu Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 2-11 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 32: Main Display Icons

    Page Down soft keys to page through the list. MORE ARROW Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 2-12 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 33: Label Location

    Figure 2-8 shows the location of the pump’s front labels, while Figure 2-9 shows the rear labels. If any labels are missing or damaged, contact your local Baxter Service Center for replacement information. MAINTAIN BATTERY CHARGE (OPTIONAL) TUBING GUIDE LABEL...
  • Page 34 UL/CSA (US ONLY) ACCESSORIES PATENTS (ON UNDERSIDE) Figure 2-9 Rear View Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 2-14 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 35: Chapter 3 Basic Operation

    Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 36: Mounting The Pump On An Iv Pole

    Turn the mounting clamp knob clockwise to close the clamp arm. Tighten until secure. Figure 3-1 Pole Mount Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 37: Recommended Administration Sets

    There are risks associated with using anything other than the recommended administration sets with this device. In the United States U.S.: Use only Baxter standard administration sets equipped with keyed On/Off ! WARNING ! Clamps that have an "s" after the code number, or are labeled as pump compatible.
  • Page 38: Preparing The Primary Infusion Container And Set

    Prepare the primary infusion container following the manufacturer’s directions for use. Attach an appropriate Baxter administration set to the solution container and prime the administration set following the directions for use. Ensure all air is expelled from the administration set.
  • Page 39: Powering On The Pump

    Figure 3-4 Pump Module Display Audible speaker is not heard during self-diagnostic test. Figure 3-5 LEDs Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 40 Feature Sets,” 5-25 for additional information. During Self-Test Figure 3-7 Power On Screen After Self-Test Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 41: Powering Off The Pump

    Tube Release,” 3-11. Figure 3-8 Main Display Screen Powering Off the Pump While the Colleague 3 Pump automatically closes the keyed On/Off Clamp, ! WARNING ! always close the regulating clamp on the administration set before removing the administration set from the pump. See “Unloading the Administration Set,”...
  • Page 42: Loading The Administration Set

    If flow is still observed, remove the pump from service and contact a Baxter authorized service person. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 43: Unloading The Administration Set

    Ensure the tubing channel is completely open (arrow displayed on the pump module display) before performing step 4. While the Colleague 3 Pump automatically closes the keyed On/Off Clamp, ! WARNING ! always close the regulating clamp on the administration set before removing the administration set from the pump.
  • Page 44: Using The Manual Tube Release

    Figure 3-11 Using the Manual Tube Release Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 3-10 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 45: Resetting The Manual Tube Release

    Manual Tube Release is reset and the pump is powered off and back on. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 3-11 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 46: Programming A Primary Rate - Volume Infusion

    Program the desired flow rate (in mL/hr) using the keypad (Figure 3-14). Figure 3-14 Rate-Volume Programming Screen Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 3-12 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 47 Programming values for each field are also stored when the pump is in stop mode and then powered off. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 3-13 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 48: Completing Or Stopping A Primary Infusion

    Channel Select to access the programming screen, then press the START key. 3-14 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724 Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT.
  • Page 49: Programming A Piggyback Infusion

    KVO alert mode after the piggyback infusion has been completed. Figure 3-19 Select Callback Yes or No Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 3-15 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 50: Preparing A Piggyback Infusion Set

    Piggyback soft key (Figure 3-20). Figure 3-20 Select Piggyback Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 3-16 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 51: Stopping A Piggyback Infusion Before Completion

    (Figure 3-22). Figure 3-22 Piggyback Stopped Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 3-17 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 52: Changing The Flow Rate During An Infusion

    Enter a new value using the keypad. Figure 3-24 Enter New Rate Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 3-18 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 53: Volume History

    Last Volumes Cleared fields. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 3-19 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 54: Using And Clearing Volume History

    However, the last date, time and total volume cleared will still be Last Volumes Cleared displayed in the field. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 3-20 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 55: Panel Lockout

    Locked Pop-up is displayed (Figure 3-28). Figure 3-28 Front Panel Locked Invalid Key Pressed Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 3-21 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 56 Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT.
  • Page 57: Chapter 4 Using The Options Menu

    Figure 4-2. Figure 4-1 Main Display Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 58: Using Flow Check Display

    Figure 4-3 Options Menu Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 59: Viewing Current Pump Personality Feature Set

    Figure 4-5 Highlight Name of Current Personality Feature Set Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 60: Overriding The Downstream Occlusion Values

    Figure 4-7 Downstream Occlusion Values Pop-Up Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 61: Auto Restart

    Use the ר Select soft key to access it. Figure 4-8 Options Menu Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 62: Using The Configuration/Service Function

    Colleague Global Service Manual for details. Figure 4-10 Options Menu Select Configuration/Service Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 63: Chapter 5 Advanced Features

    Additional programming functions and utilities ■ Using optional accessories ■ Pump Personality Feature Sets ■ Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 64: Label Library

    See “Selecting a Label,” 5-3 for selecting labels for display. Figure 5-2 Main Display Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 65: Selecting A Label

    Confirm that the selected label is appropriate for the medica- tion/solution infusing on that channel. Figure 5-4 Label List Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 66: Optional Dose Programming Functions

    Drug Amount and Diluent Volume fields are cleared. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 67: Programming A Dose Independent Of Patient Parameters

    (Figure 5-5) to display the available modes (Figure 5-6). Figure 5-5 Primary Programming Screen Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 68 To clear the parameters, press the Clear Settings soft key (Figure 5-7). Figure 5-7 Clear Old Settings Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 69 (Figure 5-9) and use the Ø field. Figure 5-9 Enter Drug Amount Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 70 The pump will then calculate the Dose. Figure 5-11 Value “HIGH” Relieved Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 71 See “Label Library,” 5-2 for details. Figure 5-13 Enter Volume to be Infused Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 72 (Figure 5-15). Figure 5-15 Confirm Settings Prompt Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 5-10 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 73: Programming A Dose Based On Patient Weight

    The Programming Modes screen is displayed (Figure 5-18). Figure 5-17 Primary Programming Screen Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 5-11 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 74 Program each field as described in “Programming a Dose Independent of Patient Parameters,” 5-5. Figure 5-19 Clear Stored Settings Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 5-12 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 75 Channel Select soft key to return to the programming screen and change the weight. The Colleague 3 Pump is intended for use in delivering multiple infusions to a ! WARNING ! single patient. Never use the pump to deliver infusions to more than one patient simultaneously.
  • Page 76: Discontinuing A Dose Program

    Restore Settings soft key to restore the programmed values. Figure 5-23 Clear Settings Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 5-14 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 77 To use other programming functions, press the Change Mode soft key (Figure 5-24). Figure 5-24 Restore Settings Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 5-15 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 78: Using The Optional Volume-Time Programming Function

    Volume-Time or Piggyback Volume-Time , then press the Select soft key. Figure 5-25 Change Mode Screen Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 5-16 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 79 Figure 5-27 Enter Time Duration Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 5-17 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 80: Using The Optional Prime Function

    Programming screen. From the Programming screen, press the soft key. Change Mode Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 5-18 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 81 Prime Figure 5-30 Ready to Prime Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 5-19 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 82 Priming volume is limited to 30 mLs. Figure 5-31 Prime Active Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 5-20 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 83: Standby Mode

    The Colleague 3 Pump does not have power on/off keys for each channel. If the device is powered on and you want to avoid a Channel Stopped alert on an unused or pre-programmed channel, you should use the Standby mode.
  • Page 84: Enabling Standby

    The display changes to the Main Display and shows Standby where Menu the program status information is normally displayed (Figure 5-34). Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 5-22 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 85: Exiting Standby Mode

    Standby and reverts to the programming mode in effect when the pump was placed into Standby. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 5-23 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 86: Using The Optional Accessories

    Using the Optional Accessories Using the Optional Accessories This pump should be used only with Baxter accessories specified in “Using the ! WARNING ! Optional Accessories,” 5-24. There are risks associated with using anything other than the recommended accessories with this pump.
  • Page 87: Event History Download Kit

    DL Event History Download Kit (product code 2M8317, U.S. only). Personality Feature Sets Overview To meet our customers’ needs, Baxter designed the Colleague 3 Pump to be highly adaptable to each care setting. Throughout this manual, references are made to specific programming functions described as configurable options.
  • Page 88: Selecting A Pump Personality Feature Set

    Select setting. Figure 5-36 Configuration Screen Example Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 5-26 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 89: Chapter 6 Troubleshooting

    STATUS LINE Figure 6-1 Main Display Status Line Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 90: Troubleshooting Alerts

    Indication: The pump is in Advance Air mode. The Advancing Air screen is displayed (Figure 6-2). Figure 6-2 Advancing Air Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 91 (Figure 6-3). Figure 6-3 Fluid Detected While the Colleague 3 Pump automatically closes the keyed On/Off Clamp, ! WARNING ! always close the regulating clamp on the administration set before removing the administration set from the pump.
  • Page 92: Troubleshooting Other Alerts

    • Stop the channel and place it in Standby mode, or • Power off the pump. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 93: Troubleshooting Alarms

    “X”. Example: X-ADV AIR (where X = A, B, or C). Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 94: Troubleshooting Air Detected Alarms

    Advance Air screen. Figure 6-5 Advance Air Pop-Up Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 95 START key. Figure 6-7 Fluid Detected Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 96: Troubleshooting Other Alarms

    (Figure 6-8). Figure 6-8 Maximum Volume Pumped While the Colleague 3 Pump automatically closes the keyed On/Off Clamp, ! WARNING ! always close the regulating clamp on the administration set before removing the administration set from the pump.
  • Page 97 START LOADING/label channel when the START key was pressed. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 98 Select key to access the appropriate programming screen and then press the START NH\ Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 6-10 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 99: Troubleshooting Failures

    See “Acknowledging Channel Failures,” 6-12. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 6-11 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 100: Acknowledging Channel Failures

    Manual Tube Release. Figure 6-10 Channel Failure Pop-up Message Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 6-12 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 101 Alarm LED will light, and the audible tone will occur. Figure 6-11 Channel Failure Display Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 6-13 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 102: Unloading The Administration Set After Device Failure

    Troubleshooting Failures Unloading the Administration Set after Device Failure While the Colleague 3 Pump automatically closes the keyed On/Off Clamp, ! WARNING ! always close the regulating clamp on the administration set before removing the administration set from the pump.
  • Page 103: Chapter 7 Maintenance And Storage

    The pump’s design safeguards against fluid spillage into the pump module. Contact your Baxter Service Center for assistance immediately if fluid enters the tubing channel. The tubing channel should be cleaned as soon as possible by authorized service personnel to minimize potential difficulties caused by fluid pooling and drying on the mechanism.
  • Page 104: Preventive Maintenance

    If the pump cannot be cleaned using the basic methods described earlier or components are missing or damaged, discontinue use and notify the appropriate authorized service personnel. To contact Baxter for authorized service or repair, see “Service Information,” 10-2. Check Action Perform as required but recommended after every use.
  • Page 105: Battery Operation Overview

    “Warranty and Service Information,” 10-1). Battery Disposal The pump contains sealed lead acid batteries which should not be disposed of in the trash. They may be returned to your authorized Baxter Service Center marked “FOR DISPOSAL.” Printed from the Baxter Intranet.
  • Page 106: Storage

    34 days (batteries removed from pump) 1. Baxter recommends removing the batteries if the pump is to be stored for more than 90 days. Storage of the pump at temperatures greater than 24°C (75°F) will hasten battery discharge which may shorten the batteries’ overall operating life.
  • Page 107: Chapter 8 Technical Specifications

    U.S.: 2M8153 U.K.: 2M8153K Triple Channel Shuttle Volumetric Infusion Pump Administration Sets U.S.: Standard Baxter administration sets equipped with keyed On/Off Clamps Latin America: Baxter “Green Label” C96XX administration sets equipped with keyed On/Off Clamps or Standard Baxter administration sets equipped with keyed On/Off clamps EEC Countries: Baxter “Green Label”...
  • Page 108 465 mmHg (9 psig) 620 mmHg (12 psig) 775 mmHg (15 psig) Maximum Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 109 See the Colleague Global Service Manual for additional information. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-G3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 110: Recommended Practices

    Healthcare Organizations) procedure. Volumetric Accuracy of the System The Colleague 3 Volumetric Infusion Pump, using the administration sets identified in Chapter 3, maintains a volumetric accuracy with delivery errors not exceeding ±5% for any one hour period over 72 hours at 100 mL/hr.
  • Page 111: Startup Graph Description

    The startup graph was developed in accordance with IEC 60601-2-24. The startup data shown in the graph illustrates the startup performance of the Colleague 3 Pump during the first 120 minutes of operation with a sampling period of 30 seconds.
  • Page 112: How Trumpet Curves Can Be Used

    Pump 90 100 Figure 8-3 Delivery Startup, First Two Hours, 1 mL/hr Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 113 Average of 5 pumps Figure 8-6 Trumpet Graph, 72nd Hour of Delivery, 1 mL/hr Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 114: Accuracy Test Per Sub-Clause 50.4 Of Iec 60601-2-24 Part 2 At 25 Ml/Hr

    37.5 12.5 Figure 8-9 Flow Accuracy, 72nd Hour, 25 mL/hr Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 115 *At least 90% of the observed values (95% confidence) will lie below the limits shown for the indicated settings. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 116 For flow rate < 1.0 mL/hr, worst case volume is 25% of set VTBI or 1.33 mL, whichever is greater. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 8-10 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 117: Chapter 9 Configurable Options

    Personality feature set selection when the pump is turned off. Note: Nurse Call is always available and produces constant output. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 118: Configurable Options List

    Enable/Disable Clinician Override of Occlusion limits ■ Set number of Auto restarts after Downstream Occlusion ■ Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 119 Enabled Mode Unavailable (Disable) Piggyback Infusion Mode Available (Enable) Enabled Unavailable (Disable) Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 120 • 25 Microliters • 50 Microliters • 100 Microliters • 150 Microliters Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 121 Individual labels can be enabled or disabled (NO) Label Library Set Up using the soft key. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 122: Label Library

    Vecuronium Bromide VECURONI Antibiotics Cefazolin Sodium CEFAZOLI Anticoagulants Alteplase ALTEPLAS Heparin Sodium HEPARIN COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724 Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT.
  • Page 123 Tolazoline HCL TOLAZOLI Trimethaphan Camsylate TRIMETHA Labor and Delivery Magnesium Sulfate MAGNESIU 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT.
  • Page 124: Application Labels

    Maintenance Line MAINTENA Total Parenteral Nutrition Umbilical Arterial Catheter Umbilical Venous Catheter Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 125: Custom Labels

    Refer to the Colleague Global Service Manual for instructions on configuring custom labels. Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 126 Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT.
  • Page 127: Chapter 10 Warranty And Service Information

    There are risks associated with using anything other than the recommended Baxter designated IV administration sets with this device. Baxter’s warranty to repair or replace the product will be null and void if this product is used contrary to the directions for use contained in the labeling or if used with non-recommended sets.
  • Page 128: Service Information

    In the event that your pump needs to be returned for service, obtain a Return Authorization by calling 1-800-THE-PUMP or your local service center. Shipping costs for all pumps returned to Baxter shall be paid for by the customer. The pump must be packed in its original container or in another Baxter approved container that will provide adequate protection during shipment.
  • Page 129: Index

    6-10 charging the battery tube not loaded 6-10 Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. Index-1 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 130 VTBI battery charge level 2-12 stopping an infusion 5-14 Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. Index-2 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 131 5-24 volume history 2-3, 3-20 3-14 piggyback 3-17 Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. Index-3 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 132 10-2 pop-up window 2-11 predefined label list preparing piggyback infusions 3-16 Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. Index-4 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 133 6-10 Standby mode 5-21 tube misloaded alarm 6-10 enabling 5-22 Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. Index-5 07-19-F3-724 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL...
  • Page 134 5-18 warning epidural infusions summary warranty 10-1 Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. Index-6 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 07-19-F3-724...
  • Page 135 Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT.
  • Page 136 Baxter Healthcare Corporation Revision G Deerfield, IL 60015 USA © Copyright 1998-2002, Baxter Healthcare Corporation. All Rights Reserved. July 2002 Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT.