ResMed S9 V-Auto Information Manual page 6

Positive airway pressure device
Table of Contents

Advertisement

Electromagnetic
compatibility
IEC 60601-1
classification
Device technical specifications
Operating pressure range (measured at the mask): 4–20 cm H
Pressure and
2–25 cm H
flow state
Maximum single fault steady state pressure: 30 cm H
for > 6 sec; 40 cm H
Pressure measurement tolerance: ±0.5 cm H
Flow measurement tolerance: ±6 L/min or 10% of reading, whichever is greater,
at 0 to 150 L/min positive flow
Mode reading
CPAP mode
Set Pressure: 4 to 20 cm H
VAuto mode
Min EPAP: 4 cm H
S mode
IPAP: 4 to 25 cm H
Physical
Nominal dimensions (L x W x H): 153 mm x 140 mm x 86 mm
Weight: 835 g
Housing construction: Flame retardant engineering thermoplastic
Air outlet: 22 mm conical air outlet (complies with ISO 5356-1:2004)
Air filter
Hypoallergenic air filter: Acrylic and polypropylene fibers in a polypropylene carrier
Standard air filter: Polyester non-woven fiber
6
Product complies with all applicable electromagnetic compatibility requirements
(EMC) according to IEC60601-1-2, for residential, commercial and light industry
environments.
It is recommended that mobile communication devices are kept at least 1 m
away from the device.
Information regarding the electromagnetic emissions and immunity of this
ResMed device can be found on www.resmed.com, on the Products page
under Service and Support. Click on the PDF file for your language.
Class II (double insulation), Type BF
O (VPAP)
2
O - if pressure exceeded for >1 sec
2
O; Max IPAP: 25 cm H
2
O; EPAP: 2 to 25 cm H
2
O ± 4% of the measured reading
2
O
2
O; Pressure support: 0 to 10 cm H
2
O
2
O (CPAP);
2
O - if pressure exceeded
2
O
2