Do you have a question about the AirCurve10 ASV and is the answer not in the manual?
Questions and answers
William Holbert
May 16, 2025
Pressure seems lower , 4th year using ResMed Aircurve 10 ASV
S/N 23202435475
1 comments:
Mr. Anderson
May 22, 2025
Possible causes of lower pressure in a ResMed AirCurve 10 ASV after four years of use include:
1. Blower motor degradation, which may lead to incorrect pressure or flow. 2. A clogged air filter reducing airflow. 3. Blocked or obstructed mask CO₂ vents. 4. Improperly seated water chamber or loose chamber lid.
Testing during calibration can confirm blower performance.
This answer is automatically generated
Garry Randall
June 9, 2025
what are the measurements of the AirCurve 10 ASV? in inches and centimeters? Height, width and depth
ENGLISH Welcome The AirCurve ™ 10 ASV is a positive airway pressure device that belongs to the adaptive servo-ventilator category. WARNING Read this entire guide before using the device. CAUTION In the US, Federal law restricts this device to sale by or on the order of a physician. AirCurve 10 ASV indications for use The AirCurve 10 ASV device is indicated for the treatment of patients weighing more than 66 lb (30 kg) with obstructive sleep apnea (OSA), central and/or mixed apneas, or periodic breathing.
At a glance The AirCurve 10 includes the following: Device with HumidAir ™ integrated humidifier • • Water tub • Air tubing • Power supply unit Travel bag • • SD card (already inserted). Contact your care provider for a range of accessories available for use with the device including: •...
About the control panel Press to start/stop therapy. Start/Stop button Press and hold for three seconds to enter power save mode. Turn to navigate the menu and press to select an option. Dial Turn to adjust a selected option and press to save your change.
Page 5
5. Close the water tub and insert it into the side of the device. 6. Connect the free end of the air tubing firmly onto the assembled mask. See the mask user guide for detailed information. Recommended masks are available on www.resmed.com.
Starting therapy 1. Fit your mask. 2. Press Start/Stop or breathe normally if SmartStart is enabled. You will know that therapy is on when the Sleep Report screen is displayed. The pressure bar shows the inspiratory and expiratory pressures in green. The green bar will expand and contract as you breathe in and out.
Power save mode Your AirCurve 10 device records your therapy data. In order to allow it to transmit the data to your care provider, you should not unplug the device. However, you can put it into power save mode to save electricity.
To adjust the Humidity Level: 1. In My Options, turn the dial to highlight Humidity Level and then press the dial. 2. Turn the dial to adjust the humidity level and press the dial to save the change. If you continue to get a dry nose or mouth, or moisture in your mask, consider using ClimateLineAir heated air tubing.
Caring for your device It is important that you regularly clean your AirCurve 10 device to make sure you receive optimal therapy. The following sections will help you with disassembling, cleaning, checking and reassembling your device. Disassembling 1. Hold the water tub at the top and bottom, press it gently and pull it away from the device. 2.
Page 10
Checking You should regularly check the water tub, air tubing and the air filter for any damage. 1. Check the water tub: • Replace it if it is leaking or has become cracked, cloudy or pitted. • Replace it if the seal is cracked or torn. •...
Page 11
Your AirCurve 10 device may be taken on board as carry-on luggage. Medical devices do not count toward your carry-on luggage limit. You can use your AirCurve 10 device on a plane as it meets the Federal Aviation Administration (FAA) requirements. Air travel compliance letters can be downloaded and printed from www.resmed.com.
Troubleshooting If you have any problems, have a look at the following troubleshooting topics. If you are not able to fix the problem, contact your care provider or ResMed. Do not try to open the device. General troubleshooting Problem/possible cause...
Problem/possible cause Solution Air pressure in my mask seems too low (it feels like I am not getting enough air) Ramp may be in progress. Wait for air pressure to build up or turn Ramp Time off. My screen is black Backlight on the screen may have turned off.
Page 14
Device message/possible cause Solution High leak detected, connect your tubing Air tubing may not be connected properly. Make sure the air tubing is firmly connected at both ends. Mask may be fitted incorrectly. Make sure your mask is fitted correctly. See your mask user guide for fitting instructions or use the Mask Fit function to check your mask fit and seal.
• If you notice any unexplained changes in the performance of the device, if it is making unusual sounds, if the device or the power supply are dropped or mishandled, or if the enclosure is broken, discontinue use and contact your care provider or your ResMed Service Center.
Page 16
CAUTION • Use only ResMed parts and accessories with the device. Non-ResMed parts may reduce the effectiveness of the treatment and/or damage the device. • Use only vented masks recommended by ResMed or by the prescribing doctor with this device.
IEC60601-1-2:2007, for residential, commercial and light industry environments. It is recommended that mobile communication devices are kept at least 1 m away from the device. Information regarding the electromagnetic emissions and immunity of this ResMed device can be found on www.resmed.com, on the Products page under Service and Support.
Page 18
Efficiency: >98% for ~7-8 micron dust; >80% for ~0.5 micron dust Aircraft use ResMed confirms that device meets the Federal Aviation Administration (FAA) requirements (RTCA/DO-160, section 21, category M) for all phases of air travel. Wireless module Technology used:...
Page 19
Accuracy may be reduced by the presence of leaks, supplemental oxygen, tidal volumes <100 mL or minute ventilation <3 L/min. Measurement accuracy verified as per ISO 10651-1:2004 for Home Care Ventilatory Support Devices (Figure 101 and Table 101) using nominal ResMed mask vent flows.
Page 20
Pressure accuracy Maximum static pressure variation at 10 cm H O according to ISO 17510-1:2007 Standard air tubing SlimLine air tubing Without humidification ± 0.5 cm H ± 0.5 cm H With humidification ± 0.5 cm H ± 0.5 cm H Maximum dynamic pressure variation according to ISO 17510-1:2007 Device without humidification and Standard air tubing / Device with humidification and Standard air tubing Pressure (cm H...
• Portable diagnostic/screening devices This warranty is only available to the initial consumer. It is not transferable. If the product fails under conditions of normal use, ResMed will repair or replace, at its option, the defective product or any of its components.
ResMed shall not be responsible for any incidental or consequential damages claimed to have resulted from the sale, installation or use of any ResMed product. Some regions or states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation may not apply to you.
Need help?
Do you have a question about the AirCurve10 ASV and is the answer not in the manual?
Questions and answers
Pressure seems lower , 4th year using ResMed Aircurve 10 ASV S/N 23202435475
Possible causes of lower pressure in a ResMed AirCurve 10 ASV after four years of use include:
1. Blower motor degradation, which may lead to incorrect pressure or flow.
2. A clogged air filter reducing airflow.
3. Blocked or obstructed mask CO₂ vents.
4. Improperly seated water chamber or loose chamber lid.
Testing during calibration can confirm blower performance.
This answer is automatically generated
what are the measurements of the AirCurve 10 ASV? in inches and centimeters? Height, width and depth