Page 1
Please read the entire Information and Welcome Guides before using your device. Indications for use The S9 V-Auto is indicated for the treatment of obstructive sleep apnoea (OSA) in patients weighing more than 30 kg. The S9 V-Auto is intended for home and hospital use.
• skin rashes. Troubleshooting If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the device(s). Problem/possible cause Solution No display Power is not connected.
Page 3
Leak Alert is enabled. SmartStart/Stop is enabled but the device does not stop automatically when you remove your mask Incompatible mask system being Only use equipment recommended by ResMed. used. The patient is using a nasal pillows Disable SmartStart/Stop.
Page 4
Turn the humidifier setting down and empty the water from resulting in accumulation of water in the air tubing. the air tubing. Check ResMed 30/90W Power Supply Unit and fully insert the Displays error message: connector The DC plug is partially inserted into Fully insert the DC plug.
Page 5
Operating altitude: Sea level to 2,591m Storage and transport temperature: -20°C to +60°C Storage and transport humidity: 10 to 95% non-condensing ResMed confirms that the device/s meets the Federal Aviation Administration Aircraft use (FAA) requirements (RTCA/DO-160, section 21, category M) for all phases of air travel.
Page 6
It is recommended that mobile communication devices are kept at least 1 m away from the device. Information regarding the electromagnetic emissions and immunity of this ResMed device can be found on www.resmed.com, on the Products page under Service and Support. Click on the PDF file for your language. IEC 60601-1...
Page 7
Sound Pressure level (CPAP mode) DECLARED With SlimLine tube: 26 dBA with uncertainty of 2 dBA as measured DUAL-NUMBER according to EN ISO 17510-1:2009 NOISE With Standard tube: 27 dBA with uncertainty of 2 dBA as measured EMISSION according to EN ISO 17510-1:2009 VALUES in With either SlimLine or 28 dBA with uncertainty of 2 dBA as measured...
3. Tube 4. Mask Flow (maximum) at set pressures The following are measured at the end of the specified tubing: Pressure, cm H S9 V-Auto and S9 V-Auto, H5i S9 V-Auto and S9 V-Auto, H5i Standard air and Standard air...
Page 9
Maximum dynamic pressure variation according to EN ISO 17510-1:2009 Pressure (cm H 10 BPM 15 BPM 20 BPM S9 V-Auto and Standard tube without H5i / S9 V-Auto and Standard tube with H5i 0.18 / 0.18 0.30 / 0.30 0.51 / 0.51 0.21 / 0.20...
If you need information on these disposal systems, please contact your local waste administration. The crossed-bin symbol invites you to use these disposal systems. If you require information on collection and disposal of your ResMed device please contact your ResMed office, local distributor or go to www.resmed.com/environment.
Portable diagnostic/screening devices This warranty is only available to the initial consumer. It is not transferable. If the product fails under conditions of normal use, ResMed will repair or replace, at its option, the defective product or any of its components.
Page 12
ResMed shall not be responsible for any incidental or consequential damages claimed to have resulted from the sale, installation or use of any ResMed product. Some regions or states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation may not apply to you.
Page 13
• Only use the ResMed 90W or 30W power supply unit. Use the 90W power supply unit to power the system comprising the device, H5i, air tubing, DC/DC converter and battery pack. The 30W power supply unit is designed to power the device only and recommended for traveling.