1.
Only use the specified halogen lamp
The power supply unit in the microscope provides the power for the halogen lamp that serves as the
microscope light source. This unit can supply power for a halogen lamp of up to 12 V-100 W.
The
specified lamp and lamphouse must be used, otherwise, the microscope could be damaged.
Specified lamphouse:
Nikon C-LP HALOGEN 12V100W model
Specified lamp:
12V-100W LONGLIFE halogen lamp (OSRAM HLX64623 or PHILIPS
7724)
Precautions for replacing the lamp
To prevent burns, allow the lamp to cool for at least thirty minutes after turning off the power switch
before replacing it. Furthermore, to prevent electrical shock and damage to the microscope, always
turn off the power switch (flip it to the "
> side) and unplug the power cord before connecting or
disconnecting the lamphouse.
After replacing the lamp, be sure to attach the lamphouse cover securely. Never use the lamphouse
without its cover.
Precautions for operating the stage
The stage rack protrudes when the stage is moved.
When operating the focus knobs or condenser
focus knob, be careful not to graze your hand on
the end of the rack.
Do not wet the microscope
If the microscope gets wet, a short circuit may result that could damage it or make it extremely hot.
If you accidentally spill a liquid on the microscope, immediately turn off the power switch (flip it to
the "
+" side) and unplug the power cord.
Then use a dry cloth to wipe away the moisture.
If any
liquid gets inside the microscope, do not use it; instead, notify your nearest Nikon representative.
Weak electromagnetic waves
This microscope emits weak electromagnetic waves.
The accuracy of any precision electronic
equipment may be adversely affected if positioned too close. If the microscope affects TV or radio
reception, move the radio or TV further away from the microscope.
Precautions for installation
Be careful not to pinch your hands or fingers when installing the microscope.
This microscope is a precision optical instrument, and using or storing it under unsuitable conditions
may damage it or may have an adverse effect on its accuracy. See "Installation location" on p. 24 and
always use the microscope in a suitable environment.