Indicators; Controls - Welch Allyn Micropaq 406 Service Manual

Table of Contents

Advertisement

Service Manual

Indicators

Visual Alarm indicator
Green
Red
Yellow
Note
Audible Alarm indicator
Beeps to indicate a patient alarm, and beeps faster for life-threatening arrhythmia alarms.
Beeps when the equipment needs attention. This beep tone is slower than the patient
alarm tones.
Volume can be configured at Acuity as high, low, or off.
Volume can be configured at Acuity for network connection or for when communications
with the FlexNet network are interrupted.

Controls

Snapshot key
When connected to Acuity, pressing this key sends a snapshot of the patient's
numeric and waveform data to the Acuity Central Monitoring System. Depending on
how Acuity is configured, this will cause Acuity to print a 21-second snapshot (14
seconds of history, 7 seconds of data after the button is pressed) to the Acuity central
station printer.
Note
Scroll up and reset Patient Alarm Tone key
Scrolls up menus on the display.
Resets a silenced patient alarm tone.
Flashes slowly during normal operation.
Flashes during patient alarm. Remains on continuously when alarms are
silenced or suspended.
Flashes during an equipment alert and while the monitor is not connected to
the network.
Remains on continuously if the operator suspends an alert at Acuity for 90
seconds or acknowledges (dismisses) a low battery alert from the monitor or
Acuity.
Flashing green indicates that the monitor is connected to the network but not
necessarily connected to a patient. If the monitor is actively monitoring a
patient, continuous green indicates that no alarms or alerts are detected.
Snapshot is the default selection of the monitor. However, the connected monitor
will inherit the configuration previously defined by Acuity. For example, if Acuity
has defined the Snapshot key to respond with a Nurse Call function and a new
monitor is introduced to the system, the Snapshot key definition will remain as
Nurse Call.
For more information about using the Acuity Central Monitoring System, refer to
Acuity Directions For Use (810-1605-XX)