Auxiliary Lamp Replacement - Scan Optics SO-5000 User Manual

Ophthalmic microscope
Table of Contents

Advertisement

SO-5000 Ophthalmic Microscope

AUXILIARY LAMP REPLACEMENT

1.
NOTE: The auxiliary light assembly and lamp may be extreme ly hot after extended
use. Use a cloth or protective glove to protect hands from the hot surfaces while
undertaking lamp maintenance or replacement if the lamp has been on recently.
2.
In the event of auxiliary lamp failure, first turn the auxiliary light off at the power supply.
4.
Remove the protective filters by unscrewing the black ring at the front of the lamp barrel.
Refer to figure 12.
5.
Push the cable which feeds into the hole in the back of the barrel until the lamp is exposed.
Remove the lamp by pulling it out of the socket. Replace the lamp with one of the same size
and rating.
6.
Replace the filter ring by screwing it back into place.
7.
Switch the auxiliary light back on at the power supply.
ISSUE NUMBER:
5.0
DATE:
11/04/02
User Manual
Figure 12
Changing the auxiliary lamp
SUPERSEDES:
4.0
DATE:
22/05/01
WRITTEN BY:
RJK
DATE:
22/08/97
Page 22 of 36