SO-5000 Ophthalmic Microscope
8.
Remove the microscope eyepiece blanks and insert the eyepieces from the tool kit. Refer to
figure 5.
M
M
9.
Connect the zoom power cable to the socket on the side of the zoom drive cover.
10..
Remove the focus control knob cover and (small) zoom control knob cover from the bag inside
the case. Refer to figure 5. Push them into position on to the knobs on the left side of the
microscope head.
11.
Remove the guide handle cover from the bag inside the case, and slide it over the guide handle
located at the front of the microscope. Refer to figure 5.
M
M
Figure 5
ISSUE NUMBER:
5.0
DATE:
11/04/02
Retain the eyepiece blanks in the tool kit for repacking the microscope. Do
not discard the eyepiece blanks.
Take care to protect the Lamphouse prism at all times. If placing the
Microscope Assembly on a bench, lie carefully on one side.
The guide handle may be used to manoeuvre the microscope during surgery.
All covers are intended to be sterilised before any operating procedures.
Inserting the eyepieces and sterilisable covers