Use in safety instrumented systems
28
End of test sequence C
• The device has failed the proof test if the expected current value/relay resistance
value at a specific level deviates by >±2 %. For troubleshooting, refer to the
Operating Instructions→ 8. This test detects 95 % dangerous, undetected failures
(proof test coverage (PTC) = 0.95).
• If the "Expert" menu group is selected, a prompt for the access code appears on the
display. If an access code has been defined under Setup → Advanced setup
→ Administration → Define access code this must be entered here. If no access code
was defined, the prompt can be acknowledged by pressing the "E" key.
Proserve NMS80
Endress+Hauser
Need help?
Do you have a question about the Proserve NMS80 and is the answer not in the manual?
Questions and answers