Use in safety instrumented systems
20
for proof testing as per test sequence A →  22 or test sequence B →  24 or test
sequence C →  26.
CAUTION
L
Unlocking the device deactivates diagnostic functions, and the device may not be able
to carry out its safety function when unlocked.
Therefore, independent measures must be taken to ensure that there is no risk of
danger while the device is unlocked.
To unlock, proceed as follows:
1. Check the position of the hardware write protection switch (dip switch marked "WP"
on the main electronics), and set this switch to "OFF".
2. Select the sequence Setup → Advanced setup → Deactivate SIL/WHG and enter the
relevant unlocking code (SIL: 7452) for the Reset write protection parameter.
 The "End of sequence" message indicates that the device was successfully
unlocked.