Moving the device:

Empty all fluid in the bath before moving device from one
place to other place�
The device must be lifted by two persons with the upper
handles�
It can also be moved on flat surface by lifting and pushing
the front of the device� It is easy to move the device with
the help of the wheels�
The angle of inclination should never be more than 10 º at
any direction when move the device!
Note: The device must not be moved during operation�
After moving the device, you must wait at least one hour
before restarting the unit�
N
Fig. 10
Item Designation
A
On/Off button:
B
LED, set:
C
LED, cooling:
D
LED display:
E
"Menu" button:
F
LED, lock:
G
Lock button:
H
"OK / Pump" button:
I
"Pump (+)" button:
J
"Pump (-)" button:
K
"Temp" button:
L
"Temp (+)" button:
M
"Temp (-)" button:
N
Fluid level indicator:

Operator panel and display

A
B C
M
L
K
Function
Switch on/off the circulator�
The LED lights up at the same time as the set value is displayed�
The LED indicates the refrigerating function is activated�
Display the settings and actual values�
Press it once: menu option is display�
Press it once again: back to working screen�
Indicate the buttons are deactivated�
Lock/unlock buttons�
Start/stop the pump function�
Confirm the menu options�
Increase the pump speed setting�
Change the menu value setting�
Decrease the pump speed setting�
Change the menu value setting�
Start/stop the tempering function�
Increase the temperature setting�
Navigation, selecting the settings in the menu�
Decrease the temperature setting�
Navigation, selecting the settings in the menu�
Indicate the fluid level (when the fluid level is too low or too high, the bottom or the
top LED segment change into red color)�
Fig. 9
D
E
F
G
GREEN
J
I
H
max. 10 ˚