CVM201 Super Bee
BAUD [Factory default = 19,200]
This sets the baud rate for the RS485 and the RS232 serial communications. The baud rate can be set to various
values through the serial interface or via the front panel soft-keys. The parity can only be changed through the
serial interface command set. When this occurs, the current setting will be shown in the list of choices and can
be re-selected if changed.
SET LINEAR [Factory default = 0.01 VOLTS to 10 VOLTS corresponding to 1 mTORR to 1 TORR]
This will take the user to four different screens to setup the linear analog output
(See Analog Output
section).
a) Set the minimum pressure
b) Set the minimum voltage corresponding to the minimum pressure
c) Set the maximum pressure
d) Set the maximum voltage corresponding to the maximum pressure
The linear output voltage can be any value between 0.01 Vdc and 10 Vdc corresponding to pressures between
1 mTorr and 1000 Torr.
INFO
This screen shows the unit software version.
ANALOG TYPE [Factory default = NONLIN]
Select "NONLIN" for non-linear (S-Curve) or "LOG" for log-linear analog output
(See Analog Output
section).

Return to Factory Default Settings

4.3
You can reset all values to the original factory default settings by simultaneously pressing the upper left and
upper right soft-keys. The user will then be prompted "Set Factory Defaults?" Choose Yes or No.
InstruTech, Inc.
Page 13