Electrical Installation; Grounding; Electrical Connections - Instrutech Super Bee CVM201 User Manual

Table of Contents

Advertisement

Instruction Manual
3.2

Electrical Installation

3.2.1

Grounding

Be sure the vacuum gauge and the rest of your vacuum system are properly grounded to protect personnel
from shock and injury. Be aware that some vacuum fittings, especially those with O-rings when not used with
metal clamps, may not produce a good electrical connection between the gauge and the chamber it is
connected to.
3.2.2

Electrical Connections

A good recommended practice is to remove power from any cable prior to connecting or disconnecting it.
The InstruTech CVM201 will directly replace Granville-Phillips® Mini-Convectron® modules that have a 9-pin
D-sub connector (DE-9P) or 15-pin D-sub connector (DE-15P), and you can use your existing cables and
electronics. For new installations, fabricate a cable to connect to the signals/functions you want to use. Signals
and pin assignments are described below:
Connector and Pinout
PIN NUMBER
PIN DESCRIPTION
1
Relay 1 Normally Open
2
Relay 1 Normally Closed
3
Power Input (11-30 Vdc)
4
Power Ground
5
Analog Output 1 (Log-Linear 1-8V, or Non-linear Granville-Phillips® Mini-Convectron ®compatible)
6
Relay 1 Common
7
Relay Disable (Disables both Relays when connected to pin 4 - Ground)
8
Analog Ground
9
Analog Output 2 (Programmable Linear 0-10 V)
PIN NUMBER
PIN DESCRIPTION
1
RS485 DATA B (+) Input/output
2
RS485 DATA A (-) Input/output
3
Power Input (11-30 Vdc)
4
Power Ground
5
Analog Output 1 (Log-Linear 1-8 V, or Non-linear Granville-Phillips® Mini-Convectron ®compatible)
6
Analog Ground
7
RS232 TX
8
RS232 RX
9
Relay Disable (Disables both Relays when connected to pin 4 - Ground)
10
Relay 1 Normally Open
11
Relay 2 Normally Open
12
Relay 2 Common
13
Relay 2 Normally Closed
14
Relay 1 Normally Closed
15
Relay 1 Common
InstruTech, Inc.
CVM201 Super Bee
Page 10

Advertisement

Table of Contents
loading

Table of Contents