Single- and multi-stream colorimetric analyzers
Note. 1
If 'New Configuration' or 'Open Configuration' is selected and the modified configuration file is saved, new data files for all
log files are created and any unarchived data is lost
Existing security configuration parameters are retained when a configuration is opened from file or when a new
configuration is loaded (the security remains as currently configured). Check 'Load security configuration from file' to
overwrite the current configuration with data from the file to be loaded.
The option to load or retain the security configuration applies only to Advanced Security mode and is available only to the
System Administrator (User 1 – see Section 6.1.4, page 28). If a new or existing configuration file is opened by a user other
than the System Administrator, existing security settings are retained.
Exiting Configuration Level
When exiting Configuration Level, the following conditions apply:
Note. 2
The current, active configuration is saved to internal storage.
Selecting 'Save as Current Configuration' suspends recording for a short time while the new configuration is implemented.
When saving the current configuration to internal storage, the file is saved automatically with a '<time><date><instrument
tag>.cfg' filename.
When saving the current configuration to external storage, the file is saved automatically to internal storage, as well as to
the external archive media as '<time><date><instrument tag>.cfg'.
When 'Save Configuration' is selected, the configuration file is stored as '<time><date><instrument tag>.cfg' on internal or
external storage.
Changes are saved to non-volatile memory only when one of the save options above has been selected. Any powerdown
before this results in lost configuration changes.
Selecting 'Cancel' discards unsaved changes and returns the analyzer to the 'Operate' level.
New internal data files for enabled recording channels are created if any of the following configuration parameters are
changed:
Recording channel source
Channel tag
A warning is displayed if a configuration change results in the creation of new internal data files for enabled recording
channels. Select 'Yes' to accept the configuration change. Select 'No' to cancel the configuration change.