R(t). Print out protocols of personalized tests depending on the user’s needs. Set up and control the instrument entirely by the USB serial port. for availability Megohmmeters Models 5050 & 5060 - User Manual GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 9
The Megohmmeter Models 5050 and 5060 are portable instruments that are housed in rugged casing. These megohmmeters are designed to check the safety of electrical installations and equipment. Measures: Voltage. Insulation Resistance. Capacitance. Leakage Current. Advantages: Automatic voltage detection in all functions, before or during the measurement, which inhibits or stops the measurement.
Page 11
2. AC power plug (direct operation on AC and battery recharge). 3. Back-lit liquid crystal display. 4. Serial interface. Male plug 9 pin for calibration only (Model 5050). USB port for connection to a PC (Model 5060). 5. Rotary selector switch with 7 positions: - Insulation measurement (to 2 T ).
Page 12
Indicates the battery charge at start-up. Indicates free memory space. One segment represents approximately 100 groups of saved values with each group holding approximately 50 recordings (1 OBJ and 1 TEST). Megohmmeters Models 5050 & 5060 - User Manual GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 13
This symbol will disappear upon pressing the disappears when the button is pressed again or if other function buttons button without having to press the button every time. Megohmmeters Models 5050 & 5060 - User Manual GlobalTestSupply www. .com Find Quality Products Online at:...
Page 15
SET-UP menu. The device will cause a fast discharge during which the insulation capacity is measured and measured again 1 minute after the residual current is circulating in the insulation. Megohmmeters Models 5050 & 5060 - User Manual GlobalTestSupply www.
Page 19
Alarm button to side of the display. Next, press the round yellow Start/Stop button to begin the test. The Alarm is tripped. Megohmmeters Models 5050 & 5060 - User Manual GlobalTestSupply www. .com Find Quality Products Online at:...
Page 21
Use the buttons. All parameters are immediately and permanently saved. Megohmmeters Models 5050 & 5060 - User Manual GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 22
The values shown on this table, in the “Display/main” and “Display/small” columns, are the factory default values. In case they are accidentally changed, it is possible to get them back. (see § 2.6.10). Megohmmeters Models 5050 & 5060 - User Manual GlobalTestSupply www.
Page 23
2400, 4800, 9600, or Parallel). “bAUd “ is not available for use. press the button three times: “SEt” appears in the small display. “ Press START Megohmmeters Models 5050 & 5060 - User Manual GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 25
500 V for the 500 V switch position on the rotary switch, and at 750 V for all the other positions on the rotary switch (LIM will appear for 3 seconds on the main display). Megohmmeters Models 5050 & 5060 - User Manual GlobalTestSupply www.
Page 26
If an external voltage higher than the value measurement, the measurement is stopped while the voltage is applied to the instrument digital display. Megohmmeters Models 5050 & 5060 - User Manual GlobalTestSupply www. .com Find Quality Products Online at:...
Page 27
(min. sec) last test voltage R(t) * The value of DD is displayed only one minute after the measurement stops. the circuit has been discharged. Megohmmeters Models 5050 & 5060 - User Manual GlobalTestSupply www. .com Find Quality Products Online at:...
Page 29
As soon as insulation measurement is stopped, the test circuit is automatically discharged using the instrument’s internal resistance. Connection diagram for measurement of strong insulation: Megohmmeters Models 5050 & 5060 - User Manual GlobalTestSupply www. .com Find Quality Products Online at:...
Page 30
R-DAR-PI-DD button. The instrument automatically measures the residual current circulating in the installation during an insulation measurement. It is displayed at the end of the measurement with the R-DAR-PI-DD button. Megohmmeters Models 5050 & 5060 - User Manual GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 32
When the button is pressed, the MR symbol is displayed. The small display indicates the last OBJ : TEST number to be assigned (e.g. 02 :11). TEST buttons to select the desired OBJ : TEST number. Megohmmeters Models 5050 & 5060 - User Manual GlobalTestSupply www. .com Find Quality Products Online at:...
Page 34
To stop printing, alter the position of the selector switch. Only the main results are printed out. Example: AEMC Instruments Model 5060 ® Instrument number: 000 001 Megohmmeters Models 5050 & 5060 - User Manual GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 37
For further information about using the Megohmmeter Control Panel, consult the Help system that comes with the product. Access this Help by clicking the option Help in the Control Panel’s menu bar at the top of the screen. Megohmmeters Models 5050 & 5060 - User Manual GlobalTestSupply www.
Page 38
Input Impedance *Over 500 Hz, the small display indicates “- - - -” and the main display gives only an assessment of the peak value of the measured voltage. Megohmmeters Models 5050 & 5060 - User Manual GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 39
Short-circuit Current: <1.6 mA ± 5 %. Load Current: 3 mA approx when starting measurement. Measurement Ranges: (40 to 5100) V (see graph below) Megohmmeters Models 5050 & 5060 - User Manual GlobalTestSupply www. .com Find Quality Products Online at:...
Page 40
Non-capacitive Test Load 500 V 40 s 1000 V 80 s 2500 V 90 s 16 s 5000 V 120 s Megohmmeters Models 5050 & 5060 - User Manual GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 41
Discharge time (C in µF) C x 3 s C x 4 s C x 4 s C x 7 s Megohmmeters Models 5050 & 5060 - User Manual GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 43
Recharge: (85 to 256) V / 50-60 Hz. (per NF EN 61557-2): Number of Measurements Nominal Charge 500 V 6500 1000 V 5500 2500 V 4000 5000 V 1500 Megohmmeters Models 5050 & 5060 - User Manual GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...