193-24-BK COMPATIBLE CURRENT 193-36-BK PROBES & SENSORS FOR A196-24-BK POWER QUALITY METERS MA193-10-BK MA193-14-BK MN93-BK MN193-BK MR193-BK SL261-BK SR193-BK E N G L I S H User Manual...
Page 3
Thank you for purchasing an AEMC current probe. For best results from your instrument and for your safety, read the enclosed operating instructions carefully and comply with the precautions for use. These products must be only used by qualified and trained users. WARNING, risk of DANGER! The operator must refer to these instructions whenever this danger symbol appears.
Page 4
PRECAUTIONS BEFORE USE The protection assured by the current probe can be compromised if it is used in a way that is not recommended by the manufacturer. ■ Comply with the rated maximum voltage and current, and the measurement category. Do not use the current probe on networks where the voltage or category exceeds those specified.
Table of Contents 1. INTRODUCTION ..............4 1.1 Receiving Your Shipment..............4 1.2 Ordering Information ................4 2. PRODUCT FEATURES ............5 2.1 Control Features ...................5 2.1.1 AmpFlex Models 193-24-BK, 193-36-BK & 196-24-BK ..5 ® 2.1.2 MiniFlex Model MA193-BK .............6 ® 2.1.3 AC/DC Current Probe Model J93 ..........7 2.1.4 AC Current Probe Models MN93-BK &...
2. PRODUCT FEATURES 2.1 Control Features 2.1.1 AmpFlex Models 193-24-BK, 193-36-BK & 196-24-BK ® Flexible sensor Sensor opening connector Shielded lead Custom 4-pin input connector Current Probes & Sensors for Power Quality Meters...
2.1.2 MiniFlex Model MA193-BK ® Flexible sensor Sensor opening device Shielded lead Custom 4-pin input connector Current Probes & Sensors for Power Quality Meters...
2.1.3 AC/DC Current Probe Model J93 ZERO 300 V CA T IV 600 V CA T III Jaws 4-point CA connector Safety guard Zero adjustment knob Power ON /Low battery indicator Three-position switch: ON, OFF, battery test Current Probes & Sensors for Power Quality Meters...
2.1.4 AC Current Probe Models MN93-BK & MN193-BK Jaws Protective guard Two-position range switch (MN193 only) Jaw opening lever Shielded lead Custom 4-pin input connector Current Probes & Sensors for Power Quality Meters...
2.1.5 AC Current Probe Model MR193-BK Jaws Protective guard Jaw opening lever Two-position range switch Shielded lead Zero adjustment Custom 4-pin input connector Current Probes & Sensors for Power Quality Meters...
2.1.6 AC Current Probe Model SL261 Jaws Zero adjust knob Range selection switch Battery compartment screw Battery compartment cover BNC Adapter (sold separately - Cat. #2140.40) Current Probes & Sensors for Power Quality Meters...
2.1.7 AC Current Probe Model SR193-BK Jaws Protective guard Jaw opening lever Shielded lead Custom 4-pin input connector Current Probes & Sensors for Power Quality Meters...
3. OPERATION The current probes and flexible sensors are used to measure the current flowing in a conductor or bus bar without opening the circuit. They also insulate the user from dangerous voltages in the circuit. The choice of current probe or sensor to be used depends on the amperage to be measured and the diameter of the cables or size of the bus bar.
Page 15
For Probes: ■ Press the jaw opening lever on the probe to open the jaws. ■ Clamp the probe around the conductor to be tested. For best results, center the conductor in the jaws of the clamp. For AmpFlex and MiniFlex Sensors: ®...
8333, 8336 & PEL Series 24" (610mm) (12,000A) AmpFlex 193-36-BK 200mA to 10kA ® 8333, 8336 & PEL Series 36" (910mm) (12,000A) AmpFlex A196-24-BK ® 200mA to 10kA 8435 24" (610mm) (12,000A) MiniFlex MA193 200mA to 3000A ® 8333, 8336 & PEL Series 10"...
11 x 10.4 x .98" 9.5 oz 10 ft (3m) 11.46" (290mm) 36" (910mm) (280 x 265 x 25mm) (220g) AmpFlex A196-24-BK ® 6.6 x 6.2 x .98" 7.7 oz 10 ft (3m) 7.64" (190mm) 24" (610mm) (170 x 158 x 25mm)
4.4 Safety Protection index IP 40 for the probes and IP 30 jaws open, according to IEC 60 529 ■ IP 65 for the AmpFlex according to IEC 60 529 ® ■ IK 04 according to IEC 50102 Drop test: According to IEC 61010-1 Electrical safety according to IEC 61010-2-032.
5. MAINTENANCE Use only factory specified replacement parts. AEMC will not be held responsible ® for any accident, incident, or malfunction following a repair done other than by its service center or by an approved repair center. CAUTION: Risk of electric shock. Disconnect the instrument from any source of electricity.
5.2.3 Model J93 ■ Insert a tool, not more than 3mm in diameter, in the hole in the battery compartment cover. ■ Push to unlock the battery compartment cover, then slide it off. ■ Remove it completely by hand. ■ Remove the battery and the shim from the compartment. ■...
5.3 Repair and Calibration To ensure that your instrument meets factory specifications, we recommend that it be scheduled back to our factory Service Center at one-year intervals for recalibration, or as required by other standards or internal procedures. For instrument repair and calibration: You must contact our Service Center for a Customer Service Authorization Number (CSA#).
5.5 Limited Warranty The current probes and sensors are warranted to the owner for a period of one year from the date of original purchase against defects in manufacture. This limited warranty is given by AEMC Instruments, not by the distributor from whom it was ®...